» Articles » PMID: 24266447

The Influence of Cortical Width of Sheep Bone on the Primary Stability of Dental Implants: an in Vitro Study

Overview
Publisher Begell House
Specialty Biotechnology
Date 2013 Nov 26
PMID 24266447
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To explore the relationship between primary implant stability and cortical bone thickness.

Materials And Methods: Twenty-three implants were inserted into fresh sheep bone. The cortical width of implant recipient sites was determined using cone-beam computed tomography (CBCT). The maximum insertion torque value of each implant was recorded. Implant stability was measured using an Osstell system. Spearman's correlation coefficient was used to evaluate the correlations among variables.

Results: The cortical width and Hounsfield unit mean values were 3.3 ± 1.0 (mm) and 1468.5 ± 272.3, respectively. The correlation between cortical bone thickness and Osstell value was not statistically significant (p > 0.01).

Conclusion: The cortical bone thickness evaluated using CBCT had no correlation with the primary stability of dental implants.

Citing Articles

Influence of Multiple Used Implant Drills on Their Cutting Performance and Fracture Resistance.

Alevizakos V, Mosch R, von See C Materials (Basel). 2023; 16(15).

PMID: 37569975 PMC: 10420324. DOI: 10.3390/ma16155271.


Primary implant stability in a bone model simulating clinical situations for the posterior maxilla: an in vitro study.

Han H, Lim H, Hong J, Ahn S, Han J, Shin S J Periodontal Implant Sci. 2016; 46(4):254-65.

PMID: 27588215 PMC: 5005813. DOI: 10.5051/jpis.2016.46.4.254.