Validation of a New Portable Metabolic System During an Incremental Running Test

Overview
Specialty Orthopedics
Date 2013 Oct 24
PMID 24149962
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

We tested a new portable metabolic system, the Jaeger Oxycon Mobile (OM) at a range of running speeds. Six subjects carried out, in random order, two incremental tests on a treadmill, one of them using the OM, and the other using the Jaeger Oxycon Pro (OP). There are systematic errors in the measurements of oxygen consumption (VO2) and respiratory exchange ratio (RER) with the OM. Production of CO2 (VCO2) tends to be overestimated by the OM, although the differences are not significant. Ventilation (VE) showed very similar values in both analyzers. Data of VO2 and RER were corrected with a regression equation which minimised the differences among the devices. The portable metabolic system OM makes systematic errors in measurements of VO2 and RER which can be adjusted with a regression analysis to obtain data comparable to those obtained by fixed systems. Key pointsPortable metabolic systems are frequently used to explore various physiological ventilatory variables in field tests.There are systematic errors in the measurements of oxygen consumption (VO2) and respiratory exchange ratio (RER) with the Jaeger Oxycon Mobile (OM) portable metabolic systemProduction of CO2 (VCO2) tends to be overestimated by the OMData of VO2 and RER can be corrected with a regression equationThe portable metabolic system OM makes systematic errors in measurements of VO2 and RER which can be adjusted with a regression analysis to obtain data comparable to those obtained by fixed systems.

Citing Articles

Enhancing instantaneous oxygen uptake estimation by non-linear model using cardio-pulmonary physiological and motion signals.

Wang Z, Zhang Q, Lan K, Yang Z, Gao X, Wu A Front Physiol. 2022; 13:897412.

PMID: 36105296 PMC: 9465676. DOI: 10.3389/fphys.2022.897412.


Telemonitoring Techniques for Lung Volume Measurement: Accuracy, Artifacts and Effort.

Mannee D, de Jongh F, van Helvoort H Front Digit Health. 2021; 2:559483.

PMID: 34713036 PMC: 8521879. DOI: 10.3389/fdgth.2020.559483.


Comprehensive use of cardiopulmonary exercise testing in pediatrics.

Woloszyn F, Mazur A Pediatr Endocrinol Diabetes Metab. 2021; 27(1):42-46.

PMID: 33599436 PMC: 10227478. DOI: 10.5114/pedm.2020.103057.


Development and validation of models to predict personal ventilation rate for air pollution research.

Good N, Carpenter T, Anderson G, Wilson A, Peel J, Browning R J Expo Sci Environ Epidemiol. 2018; 29(4):568-577.

PMID: 30185945 PMC: 6401349. DOI: 10.1038/s41370-018-0067-4.


Cardiovascular Fitness and Energy Expenditure Response during a Combined Aerobic and Circuit Weight Training Protocol.

Benito P, Alvarez-Sanchez M, Diaz V, Morencos E, Peinado A, Cupeiro R PLoS One. 2016; 11(11):e0164349.

PMID: 27832062 PMC: 5104360. DOI: 10.1371/journal.pone.0164349.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Crandall C, Taylor S, Raven P . Evaluation of the Cosmed K2 portable telemetric oxygen uptake analyzer. Med Sci Sports Exerc. 1994; 26(1):108-11. View

3.
Miller M, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A . Standardisation of spirometry. Eur Respir J. 2005; 26(2):319-38. DOI: 10.1183/09031936.05.00034805. View

4.
Macfarlane D . Automated metabolic gas analysis systems: a review. Sports Med. 2001; 31(12):841-61. DOI: 10.2165/00007256-200131120-00002. View

5.
Rietjens G, Kuipers H, Kester A, Keizer H . Validation of a computerized metabolic measurement system (Oxycon-Pro) during low and high intensity exercise. Int J Sports Med. 2001; 22(4):291-4. DOI: 10.1055/s-2001-14342. View