Comparison of Methods of Acquiring Attenuation Maps for Cardiac SPECT in the Presence of Respiratory Motion

Overview
Journal J Nucl Cardiol
Date 2013 Oct 23
PMID 24146161
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: We investigated the effect of the respiratory motion on attenuation-corrected (AC) SPECT images for three different SPECT systems, each using a different approach in obtaining attenuation maps: scanning-line sources (SLS) acquired simultaneous with emission; slow cone-beam CT (CBCT) acquired sequentially to emission; and fast helical CT (HCT) acquired sequentially to emission.

Methods: A torso phantom filled with (99m)Tc was used to model a cardiac perfusion study. Stationary baseline acquisitions were followed by the acquisitions with the phantom moving axially using a computer-controlled motion platform to simulate breathing. In addition, HCT acquisitions were made simulating breath-hold at different extents of misalignment between CT and emission. HCT images were also used to simulate the Average-CT method. Acquisitions were repeated with added breast attachments, and the heart insert in two different orientations. Visual comparison was made of AC maps, AC emission slices and polar maps. Quantitative comparisons were made of global uniformity based on the percent fractional standard deviation (%FSD) of the polar map segment values, and the ratio of the segment values in the Anterior and Inferior walls divided by that of the Lateral and Septal walls (AI/LS ratio).

Results: The AC maps for the SLS were inferior to the CT's, and most impacted by added large breast attachment. Motion artifacts seen on CBCT slices were minimized in the derived attenuation maps. AC maps obtained from HCT showed inconsistent organ sizes depending on the direction of respiration at the time of acquisition. Both visually and quantitatively CBCT resulted in the best uniformity (up to 3.4 % lower in %FSD) for all the stationary acquisitions, and for the motion acquisition of the female phantom with large breast attachment (up to 4.0 % lower). For the motion acquisition of the male phantoms, HCT resulted in slightly better uniformity (<0.5 % lower) than CBCT. Breath-hold at end-expiration slightly improved (up to 1.1 %) the uniformity over the HCT acquired during regular breathing. Further improvement was achieved with the Average-CT method. For all the systems, phantom respiratory motion reduced the AI/LS ratio compared to when the phantoms were stationary.

Conclusions: The CBCT approach resulted in the best uniformity of the AC emission images. For the female phantom with larger breast attachment, HCT and SLS were truncated at some projection angles introducing artifacts into the AC emission images. The emission image artifacts observed with HCT could be mitigated by performing breath-hold acquisition at end-expiration or Average-CT type acquisitions.

Citing Articles

Personalized Models for Injected Activity Levels in SPECT Myocardial Perfusion Imaging.

Juan Ramon A, Yang Y, Pretorius P, Johnson K, King M, Wernick M IEEE Trans Med Imaging. 2018; 38(6):1466-1476.

PMID: 30530358 PMC: 6582653. DOI: 10.1109/TMI.2018.2885319.


Investigation of dose reduction in cardiac perfusion SPECT via optimization and choice of the image reconstruction strategy.

Juan Ramon A, Yang Y, Pretorius P, Slomka P, Johnson K, King M J Nucl Cardiol. 2017; 25(6):2117-2128.

PMID: 28537039 PMC: 9407649. DOI: 10.1007/s12350-017-0920-1.


Correction of hysteretic respiratory motion in SPECT myocardial perfusion imaging: Simulation and patient studies.

Dasari P, Konik A, Pretorius P, Johnson K, Segars W, Shazeeb M Med Phys. 2016; 44(2):437-450.

PMID: 28032913 PMC: 5344194. DOI: 10.1002/mp.12072.


Differences in polar-map patterns using the novel technologies for myocardial perfusion imaging.

Zoccarato O, Marcassa C, Lizio D, Leva L, Lucignani G, Savi A J Nucl Cardiol. 2016; 24(5):1626-1636.

PMID: 27233252 DOI: 10.1007/s12350-016-0500-9.


Have SPECT artifacts breathed their last?.

Hansen C J Nucl Cardiol. 2016; 24(2):708-710.

PMID: 26892252 DOI: 10.1007/s12350-016-0428-0.


References
1.
Liu C, Xu J, Tsui B . Myocardial perfusion SPECT using a rotating multi-segment slant-hole collimator. Med Phys. 2010; 37(4):1610-8. PMC: 2852444. DOI: 10.1118/1.3310386. View

2.
Celler A, Shcherbinin S, Hughes T . An investigation of potential sources of artifacts in SPECT-CT myocardial perfusion studies. J Nucl Cardiol. 2009; 17(2):232-46. DOI: 10.1007/s12350-009-9171-0. View

3.
Pan T, Mawlawi O, Nehmeh S, Erdi Y, Luo D, Liu H . Attenuation correction of PET images with respiration-averaged CT images in PET/CT. J Nucl Med. 2005; 46(9):1481-7. View

4.
Pan T, Mawlawi O, Luo D, Liu H, Chi P, Mar M . Attenuation correction of PET cardiac data with low-dose average CT in PET/CT. Med Phys. 2006; 33(10):3931-8. DOI: 10.1118/1.2349843. View

5.
Bacharach S, Buvat I . Attenuation correction in cardiac positron emission tomography and single-photon emission computed tomography. J Nucl Cardiol. 1995; 2(3):246-55. DOI: 10.1016/s1071-3581(05)80062-5. View