Accuracy of Dual-energy Computed Tomography for the Measurement of Iodine Concentration Using Cardiac CT Protocols: Validation in a Phantom Model

Overview
Journal Eur Radiol
Specialty Radiology
Date 2013 Oct 22
PMID 24141716
Citations 33
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the accuracy of dual-energy CT (DECT) for the quantification of iodine concentrations in a thoracic phantom across various cardiac DECT protocols and simulated patient sizes.

Materials And Methods: Experiments were performed on first- and second-generation dual-source CT (DSCT) systems in DECT mode using various cardiac DECT protocols. An anthropomorphic thoracic phantom was equipped with tubular inserts containing known iodine concentrations (0-20 mg/mL) in the cardiac chamber and up to two fat-equivalent rings to simulate different patient sizes. DECT-derived iodine concentrations were measured using dedicated software and compared to true concentrations. General linear regression models were used to identify predictors of measurement accuracy

Results: Correlation between measured and true iodine concentrations (n = 72) across CT systems and protocols was excellent (R = 0.994-0.997, P < 0.0001). Mean measurement errors were 3.0 ± 7.0% and -2.9 ± 3.8% for first- and second-generation DSCT, respectively. This error increased with simulated patient size. The second-generation DSCT showed the most stable measurements across a wide range of iodine concentrations and simulated patient sizes.

Conclusion: Overall, DECT provides accurate measurements of iodine concentrations across cardiac CT protocols, strengthening the case for DECT-derived blood volume estimates as a surrogate of myocardial blood supply.

Key Points: • Dual-energy CT provides new opportunities for quantitative assessment in cardiac imaging. • DECT can quantify myocardial iodine as a surrogate for myocardial perfusion. • DECT measurements of iodine concentrations are overall very accurate. • The accuracy of such measurements decreases as patient size increases.

Citing Articles

Dual-energy Computed Tomography in Coronary Artery Calcium Scoring.

Abhijith S, Jaseemudheen M Heart Views. 2024; 25(2):74-79.

PMID: 39619105 PMC: 11604197. DOI: 10.4103/heartviews.heartviews_24_24.


Standardization of Dual-Energy CT Iodine Uptake of the Abdomen and Pelvis: Defining Reference Values in a Big Data Cohort.

Yel I, Booz C, DAngelo T, Koch V, Gruenewald L, Eichler K Diagnostics (Basel). 2024; 14(18).

PMID: 39335730 PMC: 11431114. DOI: 10.3390/diagnostics14182051.


Spectral Photon-Counting Computed Tomography: Technical Principles and Applications in the Assessment of Cardiovascular Diseases.

Meloni A, Maffei E, Clemente A, De Gori C, Occhipinti M, Positano V J Clin Med. 2024; 13(8).

PMID: 38673632 PMC: 11051476. DOI: 10.3390/jcm13082359.


Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification.

Vrbaski S, Bache S, Rajagopal J, Samei E Med Phys. 2023; 50(9):5421-5433.

PMID: 37415402 PMC: 10897956. DOI: 10.1002/mp.16583.


Influence of sex, age and thyroid function indices on dual-energy computed tomography-derived quantitative parameters of thyroid in patients with or without Hashimoto's thyroiditis.

Geng D, Zhou Y, Su G, Si Y, Shen M, Xu X BMC Med Imaging. 2023; 23(1):25.

PMID: 36740672 PMC: 9901076. DOI: 10.1186/s12880-023-00983-x.


References
1.
De Cecco C, Darnell A, Rengo M, Muscogiuri G, Bellini D, Ayuso C . Dual-energy CT: oncologic applications. AJR Am J Roentgenol. 2012; 199(5 Suppl):S98-S105. DOI: 10.2214/AJR.12.9207. View

2.
Meinel F, Graef A, Bamberg F, Thieme S, Schwarz F, Sommer W . Effectiveness of automated quantification of pulmonary perfused blood volume using dual-energy CTPA for the severity assessment of acute pulmonary embolism. Invest Radiol. 2013; 48(8):563-9. DOI: 10.1097/RLI.0b013e3182879482. View

3.
Wang R, Yu W, Wang Y, He Y, Yang L, Bi T . Incremental value of dual-energy CT to coronary CT angiography for the detection of significant coronary stenosis: comparison with quantitative coronary angiography and single photon emission computed tomography. Int J Cardiovasc Imaging. 2011; 27(5):647-56. DOI: 10.1007/s10554-011-9881-7. View

4.
Ravenel J, Leue W, Nietert P, Miller J, Taylor K, Silvestri G . Pulmonary nodule volume: effects of reconstruction parameters on automated measurements--a phantom study. Radiology. 2008; 247(2):400-8. PMC: 4148132. DOI: 10.1148/radiol.2472070868. View

5.
Vliegenthart R, Jan Pelgrim G, Ebersberger U, Rowe G, Oudkerk M, Schoepf U . Dual-energy CT of the heart. AJR Am J Roentgenol. 2012; 199(5 Suppl):S54-63. DOI: 10.2214/AJR.12.9208. View