Late Failure of Annealed Highly Cross-linked Polyethylene Acetabular Liner

Overview
Publisher Elsevier
Date 2013 Sep 3
PMID 23995342
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Highly cross-linked polyethylene (HXPE) in total hip arthroplasty (THA) has been shown to significantly decrease wear rates compared with conventional polyethylene (CPE). However, crosslinking, thermal treatment and oxidation can decrease the mechanical properties of PE, and several cases of fracture of remelted HXPE liners were reported. We present, for the first time, unexpected failures of THA with the use of annealed HXPE liners in two patients occurring at 7 and 8 years after operation. Operative findings revealed dislocated liners from the metal shell and a fracture of the superior rim at the rim-dome junction in both liners. Scanning electron microscopy showed that the cracks initiated at the rim and propagated toward the articular surface. Both liners showed generally a low amount of oxidation (less than 1.00) at the articular surface and low wear rates; however, oxidation at the rim was relatively higher (mean 1.55). These findings suggested that decreased mechanical properties at the rim-dome junction due to cross-linking, annealing and oxidation might have been caused breakage of the HXPE liners after a long implantation time, although the annealed HXPE achieved low degree of wear.

Citing Articles

Future Prospects for Clinical Applications of Nanocarbons Focusing on Carbon Nanotubes.

Saito N, Haniu H, Aoki K, Nishimura N, Uemura T Adv Sci (Weinh). 2022; 9(24):e2201214.

PMID: 35754236 PMC: 9404397. DOI: 10.1002/advs.202201214.


The Impact of Free Radical Stabilization Techniques on in vivo Mechanical Changes in Highly Cross-Linked Polyethylene Acetabular Liners.

Decker M, Price A, Khalili A, Klassen R, Walzak M, Teeter M Orthop Res Rev. 2021; 13:113-122.

PMID: 34429667 PMC: 8380133. DOI: 10.2147/ORR.S309210.


Gait instability may indicate liner failure in patients with total hip arthroplasty. A report of three cases.

Diaz-Dilernia F, Lattore M, Zanotti G, Comba F, Piccaluga F, Buttaro M Ann R Coll Surg Engl. 2021; 103(9):e298-e304.

PMID: 34414774 PMC: 9364749. DOI: 10.1308/rcsann.2021.0027.


Incarcerated fracture fragments of Longevity polyethylene liners after total hip arthroplasty.

Yee M, OKeefe T, Winter S Arthroplast Today. 2017; 2(1):6-10.

PMID: 28326389 PMC: 4957170. DOI: 10.1016/j.artd.2015.12.006.


Highly Cross-linked Polyethylene Liner Dissociation from a Cement-less Modular Acetabular Shell: Two Case Reports.

Kawano S, Sonohata M, Kitajima M, Mawatari M Open Orthop J. 2017; 10:732-740.

PMID: 28217197 PMC: 5267444. DOI: 10.2174/1874325001610010732.