» Articles » PMID: 23908066

Prospective, Multicenter, Single-arm Study of Mechanical Thrombectomy Using Solitaire Flow Restoration in Acute Ischemic Stroke

Abstract

Background And Purpose: Mechanical thrombectomy using stent retriever devices have been advocated to increase revascularization in intracranial vessel occlusion. We present the results of a large prospective study on the use of the Solitaire Flow Restoration in patients with acute ischemic stroke.

Methods: Solitaire Flow Restoration Thrombectomy for Acute Revascularization was an international, multicenter, prospective, single-arm study of Solitaire Flow Restoration thrombectomy in patients with large vessel anterior circulation strokes treated within 8 hours of symptom onset. Strict criteria for site selection were applied. The primary end point was the revascularization rate (thrombolysis in cerebral infarction ≥2b) of the occluded vessel as determined by an independent core laboratory. The secondary end point was the rate of good functional outcome (defined as 90-day modified Rankin scale, 0-2).

Results: A total of 202 patients were enrolled across 14 comprehensive stroke centers in Europe, Canada, and Australia. The median age was 72 years, 60% were female patients. The median National Institute of Health Stroke Scale was 17. Most proximal intracranial occlusion was the internal carotid artery in 18%, and the middle cerebral artery in 82%. Successful revascularization was achieved in 79.2% of patients. Device and procedure-related severe adverse events were found in 7.4%. Favorable neurological outcome was found in 57.9%. The mortality rate was 6.9%. Any intracranial hemorrhagic transformation was found in 18.8% of patients, 1.5% were symptomatic.

Conclusions: In this single-arm study, treatment with the Solitaire Flow Restoration device in intracranial anterior circulation occlusions results in high rates of revascularization, low risk of clinically relevant procedural complications, and good clinical outcomes in combination with low mortality at 90 days.

Clinical Trial Registration Url: http://www.clinicaltrials.gov. Unique identifier: NCT01327989.

Citing Articles

Iatrogenic dissection on atherosclerotic stenosis of the middle cerebral artery caused by stent retriever thrombectomy for internal carotid artery embolic occlusion.

Morishita M, Endo H, Ogino T, Fumoto K, Nakamura H Surg Neurol Int. 2025; 16:41.

PMID: 40041057 PMC: 11878664. DOI: 10.25259/SNI_928_2024.


[Complications of mechanical thrombectomy].

Mohamad J Radiologie (Heidelb). 2024; 65(2):123-126.

PMID: 39671085 DOI: 10.1007/s00117-024-01401-1.


Selective brain cooling with a novel catheter reduces infarct growth after recanalization in a canine large vessel occlusion model.

King R, Anagnostakou V, Shazeeb M, Hornibrook S, Mitchell J, Epshtein M Interv Neuroradiol. 2024; :15910199241266010.

PMID: 39043215 PMC: 11571156. DOI: 10.1177/15910199241266010.


Microfabricated Nitinol Stent Retrievers with a Micro-Patterned Surface.

Kato S, Ban Y, Ota T, Miki N Micromachines (Basel). 2024; 15(2).

PMID: 38398942 PMC: 10893500. DOI: 10.3390/mi15020213.


Diagnosis and management of occlusion in acute ischemic stroke.

Di Donna A, Muto G, Giordano F, Muto M, Guarnieri G, Servillo G Eur J Radiol Open. 2023; 11:100513.

PMID: 37609048 PMC: 10440394. DOI: 10.1016/j.ejro.2023.100513.


References
1.
Smith W, Sung G, Saver J, Budzik R, Duckwiler G, Liebeskind D . Mechanical thrombectomy for acute ischemic stroke: final results of the Multi MERCI trial. Stroke. 2008; 39(4):1205-12. DOI: 10.1161/STROKEAHA.107.497115. View

2.
Nogueira R, Lutsep H, Gupta R, Jovin T, Albers G, Walker G . Trevo versus Merci retrievers for thrombectomy revascularisation of large vessel occlusions in acute ischaemic stroke (TREVO 2): a randomised trial. Lancet. 2012; 380(9849):1231-40. PMC: 4176618. DOI: 10.1016/S0140-6736(12)61299-9. View

3.
Davalos A, Pereira V, Chapot R, Bonafe A, Andersson T, Gralla J . Retrospective multicenter study of Solitaire FR for revascularization in the treatment of acute ischemic stroke. Stroke. 2012; 43(10):2699-705. DOI: 10.1161/STROKEAHA.112.663328. View

4.
Saver J, Jahan R, Levy E, Jovin T, Baxter B, Nogueira R . Solitaire flow restoration device versus the Merci Retriever in patients with acute ischaemic stroke (SWIFT): a randomised, parallel-group, non-inferiority trial. Lancet. 2012; 380(9849):1241-9. DOI: 10.1016/S0140-6736(12)61384-1. View

5.
Ciccone A, Valvassori L, Nichelatti M, Sgoifo A, Ponzio M, Sterzi R . Endovascular treatment for acute ischemic stroke. N Engl J Med. 2013; 368(10):904-13. PMC: 3708480. DOI: 10.1056/NEJMoa1213701. View