» Articles » PMID: 23853265

CMOS Baseline Tracking and Cancellation Instrumentation for Nanoparticle-Coated Chemiresistors

Overview
Publisher IEEE
Date 2013 Jul 16
PMID 23853265
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Chemiresistor (CR) sensors and sensor arrays coated with thiolate-monolayer-protected gold nanoparticle (MPN) interfaces show great promise as detectors in gas-chromatographic microsystems with applications in biomedical and environmental analysis including breath biomarkers of disease. This paper describes a new readout circuit that overcomes the wide range of baseline resistances and drift in baseline values inherent to MPN-coated CRs to achieve a 57 ppm readout resolution. The 0.5-mum CMOS circuit operates at 5 V and provides a response resolution of 74 muV. It can cancel baseline voltages from 0.3 to 4.3 V with an accuracy of 4.2 mV and can track and compensate for drifts up to 30 mV/min. Performance was verified with MPN-coated CRs, where drift was measured and effectively cancelled. The circuit topology and size support an on-chip MPN-coated CR sensor array.

Citing Articles

A Low-Current and Multi-Channel Chemiresistor Array Sensor Device.

Wang Z, Shang G, Dinh D, Yan S, Luo J, Huang A Sensors (Basel). 2022; 22(7).

PMID: 35408393 PMC: 9003399. DOI: 10.3390/s22072781.


Towards a chemiresistive sensor-integrated electronic nose: a review.

Chiu S, Tang K Sensors (Basel). 2013; 13(10):14214-47.

PMID: 24152879 PMC: 3859118. DOI: 10.3390/s131014214.