The purpose of this study was first to evaluate the near infrared spectroscopy (NIRS) device, INVOS 4100 as a method of measuring exercise and arterial occlusion induced muscle ischemia in human leg, by comparison with InSpectra tissue spectrometer Model 325, and secondly to determine the influence of skin and subcutaneous thickness on the NIRS measurements. Twenty healthy subjects (43 ± 8 years) volunteered for the study. Tissue oxygen saturation (StO2) in the anterior tibial muscle was measured simultaneously with InSpectra Model 325 in one leg and INVOS 4100 in the contralateral leg during an exercise test until muscle exhaustion and arterial occlusion with and without exercise. The skin and subcutaneous thickness was identified by ultrasound imaging. Baseline StO2 was 87 ± 8 % detected by InSpectra and 76 ± 6 % by INVOS. Both devices detected an immediate decrease of StO2 (p < 0.001) during exercise, arterial occlusion with and without exercise, and a significant post ischemic hyperaemia (p < 0.001) during recovery. There was a significant inverse correlation between skin and subcutaneous thickness and baseline StO2 (r = -0.78, p < 0.01) as well as change in StO2 during exercise (r = -0.65, p = 0.002) for InSpectra, which was not apparent for INVOS. The results demonstrate that the cerebral/somatic oxygenation monitor (INVOS) is able to detect experimentally induced skeletal muscle ischemia in the human leg as well as the peripheral tissue spectrometer (InSpectra). Muscle oxygen saturation measurement by INVOS is less affected by skin and subcutaneous thickness than measured by InSpectra.

Citing Articles

Arterial Blood Flow and Effects on Limb Tissue Perfusion During Endoshunting of the Common Iliac Artery in an Experimental Porcine Model.

Millinger J, Langenskiold M, Nygren A, Osterberg K, Nordanstig J EJVES Vasc Forum. 2024; 61:54-61.

PMID: 38375024 PMC: 10875111. DOI: 10.1016/j.ejvsvf.2024.01.053.


Reduced oxygen desaturation in the vastus lateralis of chronic stroke survivors during graded muscle contractions.

Hyngstrom A, Nguyen J, Murphy S, Raab S, Schmit B, Gutterman D Top Stroke Rehabil. 2023; 31(5):513-526.

PMID: 38095272 PMC: 11129930. DOI: 10.1080/10749357.2023.2291902.


Non-invasive and invasive measurement of skeletal muscular oxygenation during isolated limb perfusion.

Corderfeldt Keiller A, Holmen A, Hansson C, Ricksten S, Bragadottir G, Olofsson Bagge R Perfusion. 2022; 38(5):1019-1028.

PMID: 35575302 PMC: 10265298. DOI: 10.1177/02676591221093201.


Photo-patterned oxygen sensing films based on Pt porphyrin for controlling cell growth and studying metabolism.

Zeng F, Fan Z, Wu S, Cheng X, Tian Y RSC Adv. 2022; 9(2):924-930.

PMID: 35517627 PMC: 9059522. DOI: 10.1039/c8ra09234f.


Effects of Exercise Compression Stockings on Anterior Muscle Compartment Pressure and Oxygenation During Running: A Randomized Crossover Trial Conducted in Healthy Recreational Runners.

Rennerfelt K, Lindorsson S, Brisby H, Baranto A, Zhang Q Sports Med. 2019; 49(9):1465-1473.

PMID: 31093922 PMC: 6684544. DOI: 10.1007/s40279-019-01103-y.


References
1.
Cook D, OConnor P, Eubanks S, Smith J, Lee M . Naturally occurring muscle pain during exercise: assessment and experimental evidence. Med Sci Sports Exerc. 1997; 29(8):999-1012. DOI: 10.1097/00005768-199708000-00004. View

2.
Freund P, Hobbs S, Rowell L . Cardiovascular responses to muscle ischemia in man--dependency on muscle mass. J Appl Physiol Respir Environ Exerc Physiol. 1978; 45(5):762-7. DOI: 10.1152/jappl.1978.45.5.762. View

3.
van Beekvelt M, Borghuis M, van Engelen B, Wevers R, Colier W . Adipose tissue thickness affects in vivo quantitative near-IR spectroscopy in human skeletal muscle. Clin Sci (Lond). 2001; 101(1):21-8. DOI: 10.1042/cs20000247. View

4.
Schachner T, Bonaros N, Bonatti J, Kolbitsch C . Near infrared spectroscopy for controlling the quality of distal leg perfusion in remote access cardiopulmonary bypass. Eur J Cardiothorac Surg. 2008; 34(6):1253-4. DOI: 10.1016/j.ejcts.2008.08.027. View

5.
Sako H, Hadama T, Miyamoto S, Anai H, Wada T, Iwata E . Limb ischemia and reperfusion during abdominal aortic aneurysm surgery. Surg Today. 2004; 34(10):832-6. DOI: 10.1007/s00595-004-2829-y. View