Deep Inspiration Breath-hold Technique Guided by an Opto- Electronic System for Extracranial Stereotactic Treatments

Overview
Date 2013 Jul 10
PMID 23835375
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

The purpose of this work was to evaluate the intrapatient tumor position reproducibility in a deep inspiration breath-hold (DIBH) technique based on two infrared optical tracking systems, ExacTrac and ELITETM, in stereotactic treatment of lung and liver lesions. After a feasibility study, the technique was applied to 15 patients. Each patient, provided with a real-time visual feedback of external optical marker displacements, underwent a full DIBH, a free-breathing (FB), and three consecutive DIBH CT-scans centered on the lesion to evaluate the tumor position reproducibility. The mean reproducibility of tumor position during repeated DIBH was 0.5 ± 0.3 mm in laterolateral (LL), 1.0 ± 0.9 mm in anteroposterior (AP), and 1.4 ± 0.9 mm in craniocaudal (CC) direction for lung lesions, and 1.0 ± 0.6 mm in LL, 1.1 ± 0.5 mm in AP, and 1.2 ± 0.4 mm in CC direction for liver lesions. Intra- and interbreath-hold reproducibility during treatment, as determined by optical markers displacements, was below 1 mm and 3 mm, respectively, in all directions for all patients. Optically-guided DIBH technique provides a simple noninvasive method to minimize breathing motion for collaborative patients. For each patient, it is important to ensure that the tumor position is reproducible with respect to the external markers configuration.

Citing Articles

Reproducibility of surface-based deep inspiration breath-hold technique for lung stereotactic body radiotherapy on a closed-bore gantry linac.

Nguyen D, Reinoso R, Farah J, Yossi S, Lorchel F, Passerat V Phys Imaging Radiat Oncol. 2023; 26:100448.

PMID: 37252251 PMC: 10213090. DOI: 10.1016/j.phro.2023.100448.


Intrafraction target shift comparison using two breath-hold systems in lung stereotactic body radiotherapy.

Prado A, Zucca D, de la Casa M, Marti J, Alonso L, Garcia de Acilu P Phys Imaging Radiat Oncol. 2022; 22:57-62.

PMID: 35514526 PMC: 9065403. DOI: 10.1016/j.phro.2022.04.004.


Capacitive monitoring system for real-time respiratory motion monitoring during radiation therapy.

Sadeghi P, Moran K, Robar J J Appl Clin Med Phys. 2020; 21(9):16-24.

PMID: 32643322 PMC: 7497937. DOI: 10.1002/acm2.12958.


Target position reproducibility in left-breast irradiation with deep inspiration breath-hold using multiple optical surface control points.

Fassi A, Ivaldi G, Tabarelli De Fatis P, Liotta M, Meaglia I, Porcu P J Appl Clin Med Phys. 2018; 19(4):35-43.

PMID: 29740971 PMC: 6036357. DOI: 10.1002/acm2.12321.


Recent advances in radiation oncology.

Garibaldi C, Jereczek-Fossa B, Marvaso G, Dicuonzo S, Rojas D, Cattani F Ecancermedicalscience. 2017; 11:785.

PMID: 29225692 PMC: 5718253. DOI: 10.3332/ecancer.2017.785.


References
1.
Wong J, Sharpe M, Jaffray D, Kini V, Robertson J, Stromberg J . The use of active breathing control (ABC) to reduce margin for breathing motion. Int J Radiat Oncol Biol Phys. 1999; 44(4):911-9. DOI: 10.1016/s0360-3016(99)00056-5. View

2.
Kubo H, Hill B . Respiration gated radiotherapy treatment: a technical study. Phys Med Biol. 1996; 41(1):83-91. DOI: 10.1088/0031-9155/41/1/007. View

3.
George R, Chung T, Vedam S, Ramakrishnan V, Mohan R, Weiss E . Audio-visual biofeedback for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy. Int J Radiat Oncol Biol Phys. 2006; 65(3):924-33. DOI: 10.1016/j.ijrobp.2006.02.035. View

4.
Baroni G, Garibaldi C, Riboldi M, Spadea M, Catalano G, Tagaste B . 3D optoelectronic analysis of interfractional patient setup variability in frameless extracranial stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 2005; 64(2):635-42. DOI: 10.1016/j.ijrobp.2005.07.964. View

5.
Tsunashima Y, Sakae T, Shioyama Y, Kagei K, Terunuma T, Nohtomi A . Correlation between the respiratory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy. Int J Radiat Oncol Biol Phys. 2004; 60(3):951-8. DOI: 10.1016/j.ijrobp.2004.06.026. View