Microleakage of the Prosthetic Abutment/implant Interface with Internal and External Connection: in Vitro Study

Overview
Specialty Dentistry
Date 2013 Jul 5
PMID 23822097
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: This study determined the degree of marginal microleakage of the abutment-implant interface on platforms with Morse taper connection and external connection.

Materials And Methods: For this in vitro study, 42 implants, 21 with external connection and 21 with Morse taper connection, were used, immersed in acrylic resin cylinders. Each implant was joined by a prosthetic abutment screw tightened at different degrees, forming the six study groups: (1) External connection, manual tightening (2) External connection, 20 Newton (N) tightening (3) External connection, 30 N tightening (4) Morse taper connection, manual tightening (5) Morse taper connection, 20 N tightening (6) orse taper connection, 30 N tightening. All samples were subjected to load cycling and thermocycling. Then, they were submerged in a solution of 0.2% methylene blue for 24 h. Finally, the microleakage was measured via 20× optical microscopy in each study group, average was obtained, and Mann-Whitney test was applied.

Results: Statistically significant differences (P < 0.001) were found between the levels of microleakage presented in the Morse taper connection implants (1.48) and external connection implants (2.8) in all three types of tightening. Microleakage levels decreases when increasing torque is applied to the screws.

Conclusion: Morse taper connection implants showed lower levels of microleakage than external connection implants; also, it was observed that microleakage decreases in the way torque increases.

Citing Articles

Evaluation of the influence of connection configuration on the implant-abutment interface vertical misfit of original milled titanium and laser-sintered cobalt-chromium abutments.

Diaz P, Vizoso B, Lopez-Suarez C, Gonzalo E, Mosaddad S, Suarez M Clin Oral Investig. 2025; 29(1):72.

PMID: 39841269 DOI: 10.1007/s00784-024-06093-4.


Microleakage along the implant-abutment interface: a systematic review and meta-analysis of in vitro studies.

Mao Z, Beuer F, Wu D, Zhu Q, Yassine J, Schwitalla A Int J Implant Dent. 2023; 9(1):34.

PMID: 37733145 PMC: 10514016. DOI: 10.1186/s40729-023-00494-y.


Micro-CT Evaluation of Microgaps at Implant-Abutment Connection.

Kowalski J, Puszkarz A, Radwanski M, Sokolowski J, Cichomski M, Bourgi R Materials (Basel). 2023; 16(12).

PMID: 37374674 PMC: 10305145. DOI: 10.3390/ma16124491.


Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees.

Tolesano Loureiro K, Moura M, de Oliveira G, Novais V, Zancope K, Simamoto Junior P Healthcare (Basel). 2023; 11(12).

PMID: 37372893 PMC: 10298607. DOI: 10.3390/healthcare11121775.


Influence of the Peek Abutments on Mechanical Behavior of the Internal Connections Single Implant.

de Matos J, da Rocha Scalzer Lopes G, Queiroz D, Pereira A, Sinhoreti M, Ramos N Materials (Basel). 2022; 15(22).

PMID: 36431619 PMC: 9696524. DOI: 10.3390/ma15228133.