Comparison of Heart Rate and Oxygen Saturation Measurements from Masimo and Nellcor Pulse Oximeters in Newly Born Term Infants

Overview
Journal Acta Paediatr
Specialty Pediatrics
Date 2013 Jun 27
PMID 23800004
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To compare heart rate (HR) measurements from Masimo and Nellcor pulse oximeters (POs) against HR measured via a three lead electrocardiograph (ECG) (HRECG ). We also compared peripheral oxygen saturation (SpO2 ) measurements between Nellcor and Masimo oximeters.

Method: Term infants born via elective caesarean section were studied. ECG leads were placed on the infant's chest and abdomen. Masimo and Nellcor PO sensors were randomly allocated to either foot. The monitors were placed on a trolley, and data from each monitor screen captured by a video camera. HR, SpO2 measurements and signal quality were extracted. Bland-Altman analysis was used to determine agreement between HR from the ECG and each oximeter, and between SpO2 from the oximeters.

Results: We studied 44 infants of whom 4 were resuscitated. More than 8000 pairs of observations were used for each comparison of HR and SpO2. The mean difference (±2SD) between HRECG and HRN ellcor was -0.8 (±11) beats per minute (bpm); between HRECG and HRM asimo was 0.2 (±9) bpm. The mean (±2SD) difference between SpO2Masimo and SpO2Nellcor was -3 (±15)%. The Nellcor PO measured 20% higher than the Masimo PO at SpO2 <70%.

Conclusion: Both oximeters accurately measure HR. There was good agreement between SpO2 measurements when SpO2 ≥70%. At lower SpO2 , agreement was poorer.

Citing Articles

Challenges in "probing spectroscopic probes" for noninvasive simultaneous disease diagnosis.

Roy L, Paul S, Banerjee A, Ghosh R, Mondal S, Das M Front Chem. 2025; 12():1463273.

PMID: 39844818 PMC: 11751013. DOI: 10.3389/fchem.2024.1463273.


Fast and accurate newborn heart rate monitoring at birth: A systematic review.

Kapadia V, Kawakami M, Strand M, Hurst C, Spencer A, Schmolzer G Resusc Plus. 2024; 19:100668.

PMID: 38912532 PMC: 11190559. DOI: 10.1016/j.resplu.2024.100668.


Randomised crossover study on pulse oximeter readings from different sensors in very preterm infants.

Maiwald C, Schwarz C, Bockmann K, Springer L, Poets C, Franz A Arch Dis Child Fetal Neonatal Ed. 2023; 109(4):391-396.

PMID: 38129130 PMC: 11228211. DOI: 10.1136/archdischild-2023-325961.


2023 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations: Summary From the Basic Life Support; Advanced Life Support; Pediatric Life Support; Neonatal Life Support;....

Berg K, Bray J, Ng K, Liley H, Greif R, Carlson J Circulation. 2023; 148(24):e187-e280.

PMID: 37942682 PMC: 10713008. DOI: 10.1161/CIR.0000000000001179.


Use of Pulse Oximetry during Resuscitation of 230 Newborns-A Video Analysis.

Kolstad V, Pike H, Eilevstjonn J, Buskov F, Ersdal H, Rettedal S Children (Basel). 2023; 10(7).

PMID: 37508621 PMC: 10377843. DOI: 10.3390/children10071124.