Purpose: To apply failure modes and effects analysis to high-dose-rate treatment planning to identify the most likely and significant sources of error in the process.

Methods: We have made a list of 25 failure modes grouped into six categories (imaging, catheter reconstruction, dwell position activity, dose points/normalization, optimization/dose, and evaluation). Each mode was rated on a one to five scale for severity, likelihood of occurrence, and probability of escaping detection. An overall ranking was formed from the product of the three scores. The authors assigned scores independently and the resulting rankings were averaged. We also analyzed 44 reported medical events related to high-dose-rate treatment planning listed on the Nuclear Regulatory Commission Web site and compared them with our own rankings.

Results: Failure modes associated with image sets, catheter reconstruction, indexer length, and incorrect dose points had the highest ranking in our analysis (scores higher than 20). The most often cited failure modes in the Nuclear Regulatory Commission reports examined were indexer length (20/44) and incorrect dose points (6/44). Several of our high-ranking modes are not associated with reported events.

Conclusion: It is a useful exercise to identify failure modes locally and analyze the efficacy of the local quality assurance program. Comparison with nationally reported failures can help direct the local analysis, but the absence or small number of reports for failure modes with a high score may be owing to low detectability. Such modes obviously cannot be ignored.

Citing Articles

A Novel Positive-Contrast Magnetic Resonance Imaging Line Marker for High-Dose-Rate (HDR) MRI-Assisted Radiosurgery (MARS).

Wang L, Ding Y, Bruno T, Stafford R, Lin E, Bathala T Cancers (Basel). 2024; 16(10).

PMID: 38792000 PMC: 11119838. DOI: 10.3390/cancers16101922.


Clinical implementation of failure modes and effects analysis for gynecological high-dose-rate brachytherapy.

Liu S, Jones E J Contemp Brachytherapy. 2024; 16(1):35-47.

PMID: 38584884 PMC: 10993892. DOI: 10.5114/jcb.2024.136295.


Identifying risk characteristics using failure mode and effect analysis for risk management in online magnetic resonance-guided adaptive radiation therapy.

Nishioka S, Okamoto H, Chiba T, Sakasai T, Okuma K, Kuwahara J Phys Imaging Radiat Oncol. 2022; 23:1-7.

PMID: 35712526 PMC: 9194450. DOI: 10.1016/j.phro.2022.06.002.


Quantifying clinical severity of physics errors in high-dose rate prostate brachytherapy using simulations.

Aramburu Nunez D, Trager M, Beaudry J, Cohen G, Dauer L, Gorovets D Brachytherapy. 2021; 20(5):1062-1069.

PMID: 34193362 PMC: 9283911. DOI: 10.1016/j.brachy.2021.05.007.


Strategies for effective physics plan and chart review in radiation therapy: Report of AAPM Task Group 275.

Ford E, Conroy L, Dong L, Fong de Los Santos L, Greener A, Kim G Med Phys. 2020; 47(6):e236-e272.

PMID: 31967655 PMC: 9012523. DOI: 10.1002/mp.14030.