Highly Sensitive Detection of Malaria Parasitemia in a Malaria-endemic Setting: Performance of a New Loop-mediated Isothermal Amplification Kit in a Remote Clinic in Uganda
Overview
Authors
Affiliations
Background: Current malaria diagnostic tests, including microscopy and antigen-detecting rapid tests, cannot reliably detect low-density infections. Molecular methods such as polymerase chain reaction (PCR) are highly sensitive but remain too complex for field deployment. A new commercial molecular assay based on loop-mediated isothermal amplification (LAMP) was assessed for field use.
Methods: Malaria LAMP (Eiken Chemical, Japan) was evaluated for samples from 272 outpatients at a rural Ugandan clinic and compared with expert microscopy, nested PCR, and quantitative PCR (qPCR). Two technicians performed the assay after 3 days of training, using 2 alternative blood sample-preparation methods and visual interpretation of results by fluorescence assay.
Results: Compared with 3-well nested PCR, the sensitivity of both LAMP and single-well nested PCR was 90%; the microscopy sensitivity was 51%. For samples with a Plasmodium falciparum qPCR titer of ≥ 2 parasites/µL, LAMP sensitivity was 97.8% (95% confidence interval, 93.7%-99.5%). Most false-negative LAMP results involved samples with parasitemia levels detectable by 3-well nested PCR but very low or undetectable by qPCR.
Conclusions: Malaria LAMP in a remote Ugandan clinic achieved sensitivity similar to that of single-well nested PCR in a United Kingdom reference laboratory. LAMP dramatically lowers the detection threshold achievable in malaria-endemic settings, providing a new tool for diagnosis, surveillance, and screening in elimination strategies.
Recombinase-Aided Loop-Mediated Isothermal Amplification on Human Plasmodium knowlesi.
Lai M, Abdul Hamid M, Jelip J, Mudin R, Lau Y Am J Trop Med Hyg. 2024; 110(4):648-652.
PMID: 38412548 PMC: 10993835. DOI: 10.4269/ajtmh.23-0572.
Ghosh P, Chowdhury R, Faisal K, Khan M, Hossain F, Rahat M Diagnostics (Basel). 2023; 13(24).
PMID: 38132223 PMC: 10742398. DOI: 10.3390/diagnostics13243639.
Rapid Detection of Malaria Based on Hairpin-Mediated Amplification and Lateral Flow Detection.
Zhang Y, Ke L, Sun T, Liu Y, Wei B, Du M Micromachines (Basel). 2023; 14(10).
PMID: 37893354 PMC: 10609466. DOI: 10.3390/mi14101917.
First field and laboratory evaluation of LAMP assay for malaria diagnosis in Cubal, Angola.
Febrer-Sendra B, Crego-Vicente B, Nindia A, Martinez-Camprecios J, Aixut S, Mediavilla A Parasit Vectors. 2023; 16(1):343.
PMID: 37789462 PMC: 10548721. DOI: 10.1186/s13071-023-05942-7.
Oliveira M, Caetano S, Dalot A, Sabino F, Calmeiro T, Fortunato E Analyst. 2023; 148(17):4053-4063.
PMID: 37529888 PMC: 10440799. DOI: 10.1039/d3an00971h.