Automated Indirect Immunofluorescence Antinuclear Antibody Analysis is a Standardized Alternative for Visual Microscope Interpretation

Overview
Specialties Biochemistry
Pathology
Date 2013 Apr 25
PMID 23612554
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Screening for antinuclear antibodies (ANA) is a basic tool in the serological work-up of systemic rheumatic disorders. Despite the emergence of alternative screening methods and the difficulties in standardization, indirect immunofluorescence (IIF) remains the recommended method for ANA detection. This study aimed to assess the reliability of automated ANA IIF analysis as a standardized alternative for the conventional manual approach.

Methods: ANA testing on HEp-2000 cells was performed on 304 consecutive routine sera, 28 serumbank samples displaying rare staining patterns, 219 samples of well-defined disease cohorts [141 systemic sclerosis (SSc), 13 polymyalgia rheumatica, 22 osteoarthritis, 5 ANCA-associated vasculitis and 38 spondyloarthritis] and 100 healthy donors. All samples were analyzed by automated IIF (Zenit G-sight), by conventional visual IIF microscopy and two ANA screening enzyme immunoassays (EIA).

Results: Automated and conventional ANA IIF analysis were comparable for negative/positive interpretation as well as intensity assessment (>90% agreement). In contrast, the accuracy of pattern recognition (26%) was limited. Likelihood ratios (LR) for SSc on results intervals of both Zenit G-sight and EIA increased with increasing level of positivity. Sensitivity within the SSc-associated antibody subsets was higher for Zenit G-sight (97%-100%) than EIA (10%-96%). A significant correlation between the quantitative result obtained by Zenit G-sight and the conventional end-point titer was found.

Conclusions: The use of Zenit G-sight for automated ANA IIF analysis offers opportunities to improve standardization. However, a complementary role of the expert technicians remains, especially for pattern recognition and classification of uncertain/negative samples.

Citing Articles

Evaluation of the Accuracy of Estimated Endpoint Titer of NOVA View in Indirect Immunofluorescent Antinuclear Antibody Testing.

Cho H, Jeong S, Hong J, Kim D, Park Y, Jeong S Diagnostics (Basel). 2024; 14(15).

PMID: 39125456 PMC: 11311674. DOI: 10.3390/diagnostics14151580.


Charting the Next Road Map for CSF Biomarkers in Alzheimer's Disease and Related Dementias.

Hu W, Nayyar A, Kaluzova M Neurotherapeutics. 2023; 20(4):955-974.

PMID: 37378862 PMC: 10457281. DOI: 10.1007/s13311-023-01370-8.


Clinical significance of HEp-2 cell cytoplasmic patterns in anti-neutrophil cytoplasmic antibody associated vasculitis.

Zhang N, Ji C, Bao X, Yuan C Medicine (Baltimore). 2022; 101(25):e29498.

PMID: 35758387 PMC: 9276407. DOI: 10.1097/MD.0000000000029498.


Standardization and Quality Assessment Under the Perspective of Automated Computer-Assisted HEp-2 Immunofluorescence Assay Systems.

Cinquanta L, Bizzaro N, Pesce G Front Immunol. 2021; 12:638863.

PMID: 33717188 PMC: 7947926. DOI: 10.3389/fimmu.2021.638863.


Automated Processing and Evaluation of Anti-Nuclear Antibody Indirect Immunofluorescence Testing.

Ricchiuti V, Adams J, Hardy D, Katayev A, Fleming J Front Immunol. 2018; 9:927.

PMID: 29780386 PMC: 5946161. DOI: 10.3389/fimmu.2018.00927.