Verification of Planning Target Volume Settings in Volumetric Modulated Arc Therapy for Stereotactic Body Radiation Therapy by Using In-treatment 4-dimensional Cone Beam Computed Tomography

Overview
Specialties Oncology
Radiology
Date 2013 Apr 9
PMID 23562767
Citations 24
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate setup error and tumor motion during beam delivery by using 4-dimensional cone beam computed tomography (4D CBCT) and to assess the adequacy of the planning target volume (PTV) margin for lung cancer patients undergoing volumetric modulated arc therapy for stereotactic body radiation therapy (VMAT-SBRT).

Methods And Materials: Fifteen lung cancer patients treated by single-arc VMAT-SBRT were selected in this analysis. All patients were treated with an abdominal compressor. The gross tumor volumes were contoured on maximum inspiration and maximum expiration CT datasets from 4D CT respiratory sorting and merged into internal target volumes (ITVs). The PTV margin was isotropically taken as 5 mm. Registration was automatically performed using "pre-3D" CBCT. Treatment was performed with a D95 prescription of 50 Gy delivered in 4 fractions. The 4D tumor locations during beam delivery were determined using in-treatment 4D CBCT images acquired in each fraction. Then, the discrepancy between the actual tumor location and the ITV was evaluated in the lateral, vertical, and longitudinal directions.

Results: Overall, 55 4D CBCT sets during VMAT-SBRT were successfully obtained. The amplitude of tumor motion was less than 10 mm in all directions. The average displacements between ITV and actual tumor location during treatment were 0.41 ± 0.93 mm, 0.15 ± 0.58 mm, and 0.60 ± 0.99 mm for the craniocaudal, left-right, and anteroposterior directions, respectively. The discrepancy in each phase did not exceed 5 mm in any direction.

Conclusions: With in-treatment 4D CBCT, we confirmed the required PTV margins when the registration for moving target was performed using pre-3D CBCT. In-treatment 4D CBCT is a direct method for quantitatively assessing the intrafractional location of a moving target.

Citing Articles

A review on 4D cone-beam CT (4D-CBCT) in radiation therapy: Technical advances and clinical applications.

Zhang Y, Jiang Z, Zhang Y, Ren L Med Phys. 2024; 51(8):5164-5180.

PMID: 38922912 PMC: 11321939. DOI: 10.1002/mp.17269.


Feasibility of 4D VMAT-CT.

Zhao X, Zhang R Biomed Phys Eng Express. 2022; 8(6).

PMID: 36206726 PMC: 9629170. DOI: 10.1088/2057-1976/ac9848.


Impact of four-dimensional cone-beam computed tomography on target localization for gastric mucosa-associated lymphoid tissue lymphoma radiotherapy: reducing planning target volume.

Shimohigashi Y, Toya R, Saito T, Kono Y, Doi Y, Fukugawa Y Radiat Oncol. 2021; 16(1):14.

PMID: 33446225 PMC: 7807891. DOI: 10.1186/s13014-020-01734-w.


Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study.

Kim J, Keum K, Lee H, Hong C, Park K, Kim J Radiat Oncol. 2020; 15(1):224.

PMID: 32977808 PMC: 7519557. DOI: 10.1186/s13014-020-01668-3.


ITV versus mid-ventilation for treatment planning in lung SBRT: a comparison of target coverage and PTV adequacy by using in-treatment 4D cone beam CT.

Bellec J, Arab-Ceschia F, Castelli J, Lafond C, Chajon E Radiat Oncol. 2020; 15(1):54.

PMID: 32127010 PMC: 7055092. DOI: 10.1186/s13014-020-01496-5.