» Articles » PMID: 23535316

Determination of Urinary Stone Composition Using Dual-energy Spectral CT: Initial in Vitro Analysis

Overview
Journal Clin Radiol
Specialty Radiology
Date 2013 Mar 29
PMID 23535316
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To evaluate the feasibility of determining the chemical composition of urinary stones using computed tomography (CT) spectral imaging (SI) MATERIALS AND METHODS: Two hundred and ten urinary stones were scanned in vitro using the SI mode and conventional polychromatic imaging (CPI; 120 kVp) mode. The radiodensity at 50 keV and 120 kVp, the mean calcium density (CD), and the water density (WD) were measured using the 50 keV images, 120 kVp images, calcium-based images, and water-based images, respectively. These parameters and the calcium-water ratio (CWR) were compared between different types of stones. CWR was defined as: CWR = ‌calcium density/water density‌.

Results: According to the results determined by infrared spectroscopy, 116 pure urinary stones were divided into five groups: uric acid (UA, n = 16), struvite (STR, n = 23), cystine (CYS, n = 10), calcium phosphate (CaP, n = 29), and calcium oxalate (COX, n = 38). CD, CWR, and the radiodensity of stones at 50 keV CT imaging of each group were as follows: UA (-3.55 ± 10.85 mg/ml, 0.03 ± 0.05, 510.08 ± 157.29 HU); STR (224.99 ± 64.28 mg/ml, 0.19 ± 0.07, 1058.58 ± 260.13 HU); CYS (139.67 ± 34.66 mg/ml, 0.13 ± 0.03, 725.75 ± 142.35 HU); CaP (619.39 ± 26.13 mg/ml, 0.69 ± 0.11, 2617.46 ± 186.22 HU); COX (518.61 ± 95.5 mg/ml, 0.58 ± 0.1, 2247.47 ± 379.51 HU). The differences in the CD, CRW, and radiodensity of stones at 50 keV CT imaging among the groups were statistically significant by binary comparison (p < 0.05). There were no significant differences in the radiodensity of stones at 120 keV CT imaging between UA and CYS (p = 0.267), CaP and COX (p = 0.065), and in the WD between STR and CYS (p = 0.108).

Conclusion: SI provides a novel method to better characterize pure urinary stones using the mean CD, CWR, and the radiodensity of stones at 50 keV CT imaging.

Citing Articles

Spectral performance evaluation of a second-generation spectral detector CT.

Liu L, Shapira N, Halliburton S, Meyer S, Perkins A, Litt H J Appl Clin Med Phys. 2024; 25(4):e14300.

PMID: 38386967 PMC: 11005977. DOI: 10.1002/acm2.14300.


Study on the correlation between energy spectrum computed tomography imaging and the pathological characteristics and prognosis of cervical cancer.

Pan L, Jia X, Zhao X, Zhang B, Wang S, Fan T Transl Cancer Res. 2022; 10(9):4096-4105.

PMID: 35116707 PMC: 8798028. DOI: 10.21037/tcr-21-1320.


Block matching frame based material reconstruction for spectral CT.

Wu W, Wang Q, Liu F, Zhu Y, Yu H Phys Med Biol. 2019; 64(23):235011.

PMID: 31658457 PMC: 7376577. DOI: 10.1088/1361-6560/ab51db.


Detection of gastric cancer and its histological type based on iodine concentration in spectral CT.

Li R, Li J, Wang X, Liang P, Gao J Cancer Imaging. 2018; 18(1):42.

PMID: 30413174 PMC: 6230291. DOI: 10.1186/s40644-018-0176-2.


Compositional analysis of various layers of upper urinary tract stones by infrared spectroscopy.

He Z, Jing Z, Jing-Cun Z, Chuan-Yi H, Fei G Exp Ther Med. 2017; 14(4):3165-3169.

PMID: 28912866 PMC: 5585751. DOI: 10.3892/etm.2017.4864.