Abstract

Introduction: Optimal cryoballoon ablation parameters for pulmonary vein (PV) isolation remain to be defined. We conducted a randomized preclinical trial to compare 2- versus 4-minute ablation lesions and assess the safety of active (forced) cryoballoon deflation.

Methods And Results: Thirty-two dogs underwent PV isolation with a second-generation 23 mm cryoballoon catheter. The left superior (LSPV) and inferior (LIPV) PVs were randomized in a factorial design to (1) a single 2- versus 4-minute cryoapplication, and (2) passive versus active cryoballoon deflation. Animals were survived for 30 days, after which histopathologic analysis was performed. Acute PV isolation was attained in 89.8% of PVs after a single application (93.8% LSPV, 85.2% LIPV; P = 0.2823). Mean time to PV isolation was 29.5 ± 18.5 seconds. Although 4-minute lesions were associated with a thicker neointima than 2-minute lesions (223.8 μm versus 135.6 μm; P = 0.007), no differences were observed in procedural characteristics (freezing temperature, rewarming time), rates of acute PV isolation, or the achievement of complete circumferentially transmural lesions at 30 days (78.7% overall; 86.2% for 2 minutes vs 70.0% for 4 minutes; P = 0.285). Active deflation was associated with faster balloon rewarming but not with significant differences in mean or maximum neointimal thickness.

Conclusion: A single application with the second-generation cryoballoon catheter results in a high rate of PV isolation. The degree of vascular injury was not increased by active balloon deflation and no differences in acute efficacy or mature transmural circumferential lesions were observed with 2- versus 4-minute applications.

Citing Articles

2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Tzeis S, Gerstenfeld E, Kalman J, Saad E, Sepehri Shamloo A, Andrade J J Arrhythm. 2024; 40(6):1217-1354.

PMID: 39669937 PMC: 11632303. DOI: 10.1002/joa3.13082.


Using computed tomography atrial myocardial thickness maps in cryoballoon pulmonary vein isolation: the UTMOST AF II randomized clinical trial.

Kim D, Kwon O, Hwang T, Park H, Yu H, Kim T Europace. 2024; 26(12).

PMID: 39585316 PMC: 11630070. DOI: 10.1093/europace/euae292.


2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Tzeis S, Gerstenfeld E, Kalman J, Saad E, Sepehri Shamloo A, Andrade J J Interv Card Electrophysiol. 2024; 67(5):921-1072.

PMID: 38609733 DOI: 10.1007/s10840-024-01771-5.


2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Tzeis S, Gerstenfeld E, Kalman J, Saad E, Sepehri Shamloo A, Andrade J Europace. 2024; 26(4).

PMID: 38587017 PMC: 11000153. DOI: 10.1093/europace/euae043.


Variation in the frozen lesion size according to the non-occluded application duration and technique for cryoballoon ablation.

Kawaji T, Bao B, Hojo S, Tezuka Y, Nakatsuma K, Matsuda S PLoS One. 2024; 19(1):e0297263.

PMID: 38271400 PMC: 10810503. DOI: 10.1371/journal.pone.0297263.