Optimizing 4D Cone Beam Computed Tomography Acquisition by Varying the Gantry Velocity and Projection Time Interval

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2013 Feb 23
PMID 23429168
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Four dimensional cone beam computed tomography (4DCBCT) is an emerging clinical image guidance strategy for tumour sites affected by respiratory motion. In current generation 4DCBCT techniques, both the gantry rotation speed and imaging frequency are constant and independent of the patient's breathing which can lead to projection clustering. We present a mixed integer quadratic programming (MIQP) model for respiratory motion guided-4DCBCT (RMG-4DCBCT) which regulates the gantry velocity and projection time interval, in response to the patient's respiratory signal, so that a full set of evenly spaced projections can be taken in a number of phase, or displacement, bins during the respiratory cycle. In each respiratory bin, an image can be reconstructed from the projections to give a 4D view of the patient's anatomy so that the motion of the lungs, and tumour, can be observed during the breathing cycle. A solution to the full MIQP model in a practical amount of time, 10 s, is not possible with the leading commercial MIQP solvers, so a heuristic method is presented. Using parameter settings typically used on current generation 4DCBCT systems (4 min image acquisition, 1200 projections, 10 respiratory bins) and a sinusoidal breathing trace with a 4 s period, we show that the root mean square (RMS) of the angular separation between projections with displacement binning is 2.7° using existing constant gantry speed systems and 0.6° using RMG-4DCBCT. For phase based binning the RMS is 2.7° using constant gantry speed systems and 2.5° using RMG-4DCBCT. The optimization algorithm presented is a critical step on the path to developing a system for RMG-4DCBCT.

Citing Articles

Optimising 4D imaging of fast-oscillating structures using X-ray microtomography with retrospective gating.

Klos A, Bailly L, Rolland du Roscoat S, Orgeas L, Henrich Bernardoni N, Broche L Sci Rep. 2024; 14(1):20499.

PMID: 39227377 PMC: 11372196. DOI: 10.1038/s41598-024-68684-1.


A review on 4D cone-beam CT (4D-CBCT) in radiation therapy: Technical advances and clinical applications.

Zhang Y, Jiang Z, Zhang Y, Ren L Med Phys. 2024; 51(8):5164-5180.

PMID: 38922912 PMC: 11321939. DOI: 10.1002/mp.17269.


Faster and lower dose imaging: evaluating adaptive, constant gantry velocity and angular separation in fast low-dose 4D cone beam CT imaging.

Lau B, Dillon O, Vinod S, OBrien R, Reynolds T Med Phys. 2023; 51(2):1364-1382.

PMID: 37427751 PMC: 11528889. DOI: 10.1002/mp.16585.


X-ray source arrays for volumetric imaging during radiotherapy treatment.

Dillon O, Reynolds T, OBrien R Sci Rep. 2023; 13(1):9776.

PMID: 37328551 PMC: 10275902. DOI: 10.1038/s41598-023-36708-x.


Intrafraction 4D-cone beam CT acquired during volumetric arc radiotherapy delivery: kV parameter optimization and 4D motion accuracy for lung stereotactic body radiotherapy (SBRT) patients.

Liang J, Lack D, Zhou J, Liu Q, Grills I, Yan D J Appl Clin Med Phys. 2019; 20(12):10-24.

PMID: 31675150 PMC: 6909121. DOI: 10.1002/acm2.12755.