Macro Design Effects on Stress Distribution Around Implants: a Photoelastic Stress Analysis
Overview
Authors
Affiliations
Objectives: Biomechanics is one of the main factors for achieving long-term success of implant supported prostheses. Long-term failures mostly depend on biomechanical complications. It is important to distinguish the effects of macro design of the implants.
Materials And Methods: In this study, the photoelastic response of four different types of implants that were inserted with different angulations were comparatively analyzed. The implant types investigated were screw cylinder (ITI, Straumann AG, Basel, Switzerland), stepped cylinder (Frialit2, Friadent GmbH, Manheim, Germany), root form (Camlog Rootline, Alatatec, Wilshelm, Germany), and cylindrical implant, with micro-threads on the implant neck (Astra, AstraTech, Mölndal, Sweden). In the test models, one of the implants was inserted straight, while the other one was aligned mesially with 15° angles. The superstructures were prepared as single crowns. A 150N loading was applied to the restorations throughout the test.
Results: A comparison of the implant designs showed that there were no significant differences between the straight implants; however, between the inclined implants, the most favorable stress distribution was seen with the stepped cylinder implants. The least favorable stress concentration was observed around the root formed implants. Microthreads around the implant neck appeared to be effective in a homogenous stress distribution. Observations showed that misaligned implants caused less stress than straight implants, but the stress concentrations were not homogenous.
Conclusion: As there were observable differences between the implant types, straight placed cylindrical implants showed better stress distribution characteristics, while inclined tapering implants had better stress distribution characteristics.
Effects of crown retrieval on implants and the surrounding bone: a finite element analysis.
Ozkir S, Mutluay Unal S, Yurekli E, Guven S J Adv Prosthodont. 2016; 8(2):131-6.
PMID: 27141257 PMC: 4852265. DOI: 10.4047/jap.2016.8.2.131.
Yingmin S, Guangbao S, Lingfeng H, Zhenwei Z Hua Xi Kou Qiang Yi Xue Za Zhi. 2014; 32(5):450-4.
PMID: 25490820 PMC: 7041027.
Photoelastic and finite element analyses of occlusal loads in mandibular body.
Rossi A, Freire A, Prado F, Asprino L, Correr-Sobrinho L, Caria P Anat Res Int. 2014; 2014:174028.
PMID: 25374695 PMC: 4206929. DOI: 10.1155/2014/174028.