Clarithromycin and rifampicin are used for the treatment of Mycobacteria. Pharmacokinetic drug interaction is possibly due to the influence of the two drugs on the liver enzymes. Using a Hypurity Aquastar C18 column (50mm×2.1mm×5μm) for liquid chromatography including a polar end-capped phase for the determination of clarithromycin, rifampicin and their metabolites together in plasma using LC-MS/MS resulted in a substantial carry-over. As a consequence, the throughput of the method is not assured. Using a step-by-step troubleshooting procedure, such carry-over was found originating from column memory effect. With the use of another type of C18 column, the carry-over is eliminated. Due to the absence of carry-over, the analytical concentration ranges are extended and are therefore more appropriate for the analysis of patient samples. The method was re-validated for linearity, reproducibility and dilution integrity.
A Simple UPLC-MS/MS Assay of Rifampin in a Small Volume of Human Plasma.
Sano T, Ishii T, Hotta K, Mano Y ACS Omega. 2023; 8(39):36261-36268.
PMID: 37810660 PMC: 10552088. DOI: 10.1021/acsomega.3c04814.
Xing Y, Yin L, Le X, Chen J, Zhang L, Li Y Heliyon. 2021; 7(7):e07532.
PMID: 34296020 PMC: 8282971. DOI: 10.1016/j.heliyon.2021.e07532.
Zhao F, Wei Y, Yan Y, Liu H, Zhou S, Ren B Int J Anal Chem. 2020; 2020:8837254.
PMID: 33381185 PMC: 7762669. DOI: 10.1155/2020/8837254.
Troubleshooting Carry-Over in the LC-MS Analysis of Biomolecules: The Case of Neuropeptide Y.
Yamagaki T, Yamazaki T Mass Spectrom (Tokyo). 2020; 8(2):S0083.
PMID: 33299733 PMC: 7709882. DOI: 10.5702/massspectrometry.S0083.
Fast and Simple LC-MS/MS Method for Rifampicin Quantification in Human Plasma.
Temova Rakusa Z, Roskar R, Klancar Andrejc A, Trdan Lusin T, Faganeli N, Grabnar I Int J Anal Chem. 2019; 2019:4848236.
PMID: 30853987 PMC: 6377990. DOI: 10.1155/2019/4848236.