» Articles » PMID: 23334924

Rapid Freehand MR-guided Percutaneous Needle Interventions: an Image-based Approach to Improve Workflow and Feasibility

Overview
Date 2013 Jan 22
PMID 23334924
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To develop and evaluate software-based methods for improving the workflow of magnetic resonance (MR)-guided percutaneous interventions.

Materials And Methods: A set of methods was developed that allows the user to: 1) plan an entire procedure, 2) directly apply this plan to skin entry site localization without further imaging, and 3) place a needle under real-time MR guidance with automatic alignment of three orthogonal slices along a planned trajectory with preference to the principal patient axes. To validate targeting accuracy and time, phantom experiments (96 targets) and in vivo paraspinal and kidney needle punctures in two pigs (55 targets) were performed. The influence of trajectory obliquity, level of experience, and organ motion on targeting accuracy and time was analyzed.

Results: Mean targeting error was 1.8 ± 0.9 mm (in vitro) and 2.9 ± 1.0 mm (in vivo) in all directions. No statistically significant differences in targeting accuracy between single- and double-oblique trajectories, novice and expert users, or paraspinal and kidney punctures were observed. The average time (in vivo) from trajectory planning to verification of accurate needle placement was 6 minutes.

Conclusion: The developed methods allow for accurate needle placement along complex trajectories and are anticipated to reduce table time for MR-guided percutaneous needle interventions.

Citing Articles

Magnetic resonance - guided treatment of low-flow vascular malformations and the technologies to potentiate adoption.

Bailey C, Herrera D, Neumeister N, Weiss C Front Med (Lausanne). 2024; 11:1319046.

PMID: 38420359 PMC: 10899448. DOI: 10.3389/fmed.2024.1319046.


Integration and evaluation of a gradient-based needle navigation system for percutaneous MR-guided interventions.

Pan L, Valdeig S, Kagebein U, Qing K, Fetics B, Roth A PLoS One. 2020; 15(7):e0236295.

PMID: 32706813 PMC: 7380643. DOI: 10.1371/journal.pone.0236295.


Modeling of active shimming of metallic needles for interventional MRI.

Sengupta S Magn Reson Med. 2020; 84(5):2858-2870.

PMID: 32597521 PMC: 7396316. DOI: 10.1002/mrm.28320.


Concepts for augmented reality visualisation to support needle guidance inside the MRI.

Mewes A, Heinrich F, Hensen B, Wacker F, Lawonn K, Hansen C Healthc Technol Lett. 2018; 5(5):172-176.

PMID: 30464849 PMC: 6222244. DOI: 10.1049/htl.2018.5076.


Percutaneous MR-guided interventions using an optical Moiré Phase tracking system: Initial results.

Kagebein U, Godenschweger F, Armstrong B, Rose G, Wacker F, Speck O PLoS One. 2018; 13(10):e0205394.