Experimental Characterization of a Prototype Detector System for Carbon Ion Radiography and Tomography

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2013 Jan 9
PMID 23296259
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Ion beams exhibit a finite range and an inverted depth-dose profile, the Bragg peak. These favorable physical properties allow excellent tumor-dose conformality. However, they introduce sensitivity to range uncertainties. Although these uncertainties are typically taken into account in treatment planning, delivery of the intended dose to the patient has to be ensured daily to prevent underdosage of the tumor or overdosage of surrounding critical structures. Thus, imaging techniques play an increasingly important role for treatment planning and in situ monitoring in ion beam therapy. At the Heidelberg Ion Beam Therapy (HIT) center, a prototype detector system based on a stack of 61 ionization chambers has been assembled for the purpose of radiographic and tomographic imaging of transmitted energetic ions. Its applicability to ion-based transmission imaging was investigated experimentally. An extensive characterization of the set-up in terms of beam parameters and settings of the read-out electronics was performed. Overall, the findings of this work support the potential of an efficient experimental set-up as the range telescope equipped with high sensitivity and fast electronics to perform heavy ion radiography and tomography at HIT.

Citing Articles

Helium ion beam imaging for image guided ion radiotherapy.

Martisikova M, Gehrke T, Berke S, Arico G, Jakel O Radiat Oncol. 2018; 13(1):109.

PMID: 29898746 PMC: 6000951. DOI: 10.1186/s13014-018-1046-6.


Particle therapy of moving targets-the strategies for tumour motion monitoring and moving targets irradiation.

Kubiak T Br J Radiol. 2016; 89(1066):20150275.

PMID: 27376637 PMC: 5124789. DOI: 10.1259/bjr.20150275.


High-energy proton imaging for biomedical applications.

Prall M, Durante M, Berger T, Przybyla B, Graeff C, Lang P Sci Rep. 2016; 6:27651.

PMID: 27282667 PMC: 4901340. DOI: 10.1038/srep27651.


Proton-counting radiography for proton therapy: a proof of principle using CMOS APS technology.

Poludniowski G, Allinson N, Anaxagoras T, Esposito M, Green S, Manolopoulos S Phys Med Biol. 2014; 59(11):2569-81.

PMID: 24785680 PMC: 4127315. DOI: 10.1088/0031-9155/59/11/2569.


Advances in 4D treatment planning for scanned particle beam therapy - report of dedicated workshops.

Bert C, Graeff C, Riboldi M, Nill S, Baroni G, Knopf A Technol Cancer Res Treat. 2013; 13(6):485-95.

PMID: 24354749 PMC: 4527425. DOI: 10.7785/tcrtexpress.2013.600274.