In Vivo Assessment of Stent Recoil of Biodegradable Polymer-coated Cobalt-chromium Sirolimus-eluting Coronary Stent System
Overview
Authors
Affiliations
Introduction: Immediate and acute stent recoil has been observed following balloon deflation in normal and diseased coronary arteries, and the degree varies by stent design.
Methods: A total of 19 patients, who underwent elective stent implantation for single de novo native coronary artery lesions, were enrolled: all patients treated with the biodegradable polymer-coated sirolimus-eluting cobalt-chromium coronary stent system (Supralimus-Core(®)). The immediate, acute and cumulative stent recoil was assessed by quantitative coronary angiography. The cumulative stent recoil was measured at 24 h of stent implantation.
Results: The absolute late loss due to recoil was found 0.08 ± 0.19 mm for Immediate Stent Recoil (ISR), 0.05 ± 0.21 mm for Acute Stent Recoil (ASR) and 0.11 ± 0.25 mm for Cumulative Stent Recoil (CSR) respectively.
Conclusions: In vivo acute stent recoil of the Supralimus-Core(®) has higher radial strength compared to other available standard drug-eluting stents.
Hataminia F, Faridi-Majidi R, Hashemi S, Ghanbari H Sci Rep. 2025; 15(1):7454.
PMID: 40032856 PMC: 11876349. DOI: 10.1038/s41598-025-86111-x.
Garcia-Villen F, Lopez-Zarraga F, Viseras C, Ruiz-Alonso S, Al-Hakim F, Diez-Aldama I Int J Bioprint. 2023; 9(2):664.
PMID: 37065659 PMC: 10090821. DOI: 10.18063/ijb.v9i2.664.
das Neves R, Mortari M, Schwartz E, Kipnis A, Junqueira-Kipnis A Toxins (Basel). 2019; 11(4).
PMID: 30974767 PMC: 6520840. DOI: 10.3390/toxins11040216.
Shetty R, Prajapati J, Pai U, Shetty K Scientifica (Cairo). 2016; 2016:9324279.
PMID: 27597929 PMC: 5002302. DOI: 10.1155/2016/9324279.
Sarma R, Prajapati J, Raheem A, Kamlesh Thakkar , Kothari S, Thakkar A J Clin Diagn Res. 2015; 9(8):OC23-6.
PMID: 26435986 PMC: 4576579. DOI: 10.7860/JCDR/2015/14060.6403.