Validation of the IHealth BP3 Upper-arm Blood Pressure Monitor, for Clinic Use and Self-measurement, According to the European Society of Hypertension International Protocol Revision 2010

Overview
Date 2012 Nov 14
PMID 23147535
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: This study aimed to evaluate the performance of the iHealth BP3 upper-arm blood pressure monitor, which is designed for clinic use and self-measurement of blood pressure using Apple touch devices as an interface.

Methods: The European Society of Hypertension International Protocol (ESH-IP) revision 2010 for the validation of blood pressure measuring devices in adults was followed precisely. Ninty-nine couples of test device and reference blood pressure measurements were obtained during the study (three pairs for each of the 33 participants).

Results: The 33 participants, age 47.1±12.3 years (age range 27-69 years) and arm circumference 30.0±4.4 cm, had a mean systolic blood pressure (SBP) of 143.9±27.4 mmHg and a mean diastolic blood pressure (DBP) of 90.1±18.3 mmHg. The device passed all of the requirements fulfilling the standards of the protocol, and the mean±SD device-observer difference was 2.8±4.2 mmHg for SBP and -0.4±3.5 mmHg for DBP.

Conclusion: According to the results of the validation study on the basis of the ESH-IP revision 2010, the iHealth BP3 can be recommended for clinic use and self-measurement in an adult population.