A Nonlinear QCT-based Finite Element Model Validation Study for the Human Femur Tested in Two Configurations in Vitro

Overview
Journal Bone
Date 2012 Sep 19
PMID 22985891
Citations 43
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Femoral fracture is a common medical problem in osteoporotic individuals. Bone mineral density (BMD) is the gold standard measure to evaluate fracture risk in vivo. Quantitative computed tomography (QCT)-based homogenized voxel finite element (hvFE) models have been proved to be more accurate predictors of femoral strength than BMD by adding geometrical and material properties. The aim of this study was to evaluate the ability of hvFE models in predicting femoral stiffness, strength and failure location for a large number of pairs of human femora tested in two different loading scenarios.

Methods: Thirty-six pairs of femora were scanned with QCT and total proximal BMD and BMC were evaluated. For each pair, one femur was positioned in one-legged stance configuration (STANCE) and the other in a sideways configuration (SIDE). Nonlinear hvFE models were generated from QCT images by reproducing the same loading configurations imposed in the experiments. For experiments and models, the structural properties (stiffness and ultimate load), the failure location and the motion of the femoral head were computed and compared.

Results: In both configurations, hvFE models predicted both stiffness (R(2)=0.82 for STANCE and R(2)=0.74 for SIDE) and femoral ultimate load (R(2)=0.80 for STANCE and R(2)=0.85 for SIDE) better than BMD and BMC. Moreover, the models predicted qualitatively well the failure location (66% of cases) and the motion of the femoral head.

Conclusions: The subject specific QCT-based nonlinear hvFE model cannot only predict femoral apparent mechanical properties better than densitometric measures, but can additionally provide useful qualitative information about failure location.

Citing Articles

Human fall simulation testing method: where we are.

Han S, Jiang X, Gao Y Osteoporos Int. 2024; 36(1):35-45.

PMID: 39556250 DOI: 10.1007/s00198-024-07316-w.


Computer-aided diagnosis for China-Japan Friendship Hospital classification of necrotic femurs using statistical shape and appearance model based on CT scans.

Zhang J, Gong H, Ren P, Liu S, Jia Z, Shi P Med Biol Eng Comput. 2024; 63(3):867-883.

PMID: 39538108 DOI: 10.1007/s11517-024-03239-0.


Segmentation methods for quantifying X-ray Computed Tomography based biomarkers to assess hip fracture risk: a systematic literature review.

Falcinelli C, Cheong V, Ellingsen L, Helgason B Front Bioeng Biotechnol. 2024; 12:1446829.

PMID: 39506973 PMC: 11537876. DOI: 10.3389/fbioe.2024.1446829.


Method for computer tomography voxel-based finite element analysis and validation with digital image correlation system.

Paul D, Arwood Z, Mulon P, Penumadu D, Truster T MethodsX. 2024; 13:102879.

PMID: 39206058 PMC: 11350451. DOI: 10.1016/j.mex.2024.102879.


CT-based finite element simulating spatial bone damage accumulation predicts metastatic human vertebrae strength and stiffness.

Soltani Z, Xu M, Radovitzky R, Stadelmann M, Hackney D, Alkalay R Front Bioeng Biotechnol. 2024; 12:1424553.

PMID: 39108596 PMC: 11300227. DOI: 10.3389/fbioe.2024.1424553.