Application of Immunomagnetic Particles to Enzyme-linked Immunosorbent Assay (ELISA) for Improvement of Detection Sensitivity of HCG
Overview
Affiliations
This investigation was aimed at using superparamagnetic particles to enzyme-linked immunosorbent assay (SPIO-ELISA) of human chorionic gonadotropin (hCG) to enhance detection sensitivity of hCG. We found that N-(3-dimethyl aminopropyl)-N'-ethylcarbodiimide hydrochloride (EDC) was the best cross-linking reagent to link anti hCG α antibody to superparamagnetic particle (SPIO-anti hCG α antibody immunomagnetic particle). To improve the specificity of the assay, a horse radish peroxidase (HRP)-labeled anti-hCG beta monoclonal antibody was used to detect captured hCG using double antibody sandwich ELISA assay. SPIO-ELISA application to determine hCG increased the sensitivity to 1 mIU/mL, which is a level of sensitivity enabling the diagnosis of pregnancy during the early gestational period.
Enzyme-Linked Immunosorbent Assay: Types and Applications.
Hayrapetyan H, Tran T, Tellez-Corrales E, Madiraju C Methods Mol Biol. 2023; 2612:1-17.
PMID: 36795355 DOI: 10.1007/978-1-0716-2903-1_1.
Nanotechnologies in Obstetrics and Cancer during Pregnancy: A Narrative Review.
Bertozzi S, Corradetti B, Seriau L, Diaz Nanez J, Cedolini C, Fruscalzo A J Pers Med. 2022; 12(8).
PMID: 36013273 PMC: 9410527. DOI: 10.3390/jpm12081324.
Emerging technologies to increase ligand binding assay sensitivity.
Fischer S, Joyce A, Spengler M, Yang T, Zhuang Y, Fjording M AAPS J. 2014; 17(1):93-101.
PMID: 25331105 PMC: 4287293. DOI: 10.1208/s12248-014-9682-8.