Pulse Oximetry and High-dose Vasopressors: a Comparison Between Forehead Reflectance and Finger Transmission Sensors

Overview
Specialty Critical Care
Date 2012 Aug 8
PMID 22868275
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Pulse oximetry (SpO(2)) measured at finger site via transmission mode may fail in situations of hypoperfusion. Forehead sensors using reflectance technology might be useful in these circumstances. We hypothesized that reflectance SpO(2) would be more accurate than finger SpO(2) in patients with severe shock.

Methods: A prospective observational study was conducted in an intensive care unit of a university hospital of patients in shock who were treated with high norepinephrine and/or epinephrine doses (≥0.1 μg kg(-1) min(-1)). When blood gas determinations were requested, forehead SpO(2) and finger SpO(2) values were simultaneous recorded. Agreement between SpO(2) measurements with arterial saturation (SaO(2)), obtained by blood analysis with a co-oximeter, was assessed using the Bland-Altman method. The number of outliers, defined by the formula SaO(2) - SpO(2) > ±3 %, indicated the proportion of measurements considered to be clinically unacceptable.

Results: Thirty-two patients were enrolled in the study. With the forehead sensor no reading failure occurred, and 140 paired data sets (forehead SpO(2) vs. SaO(2)) were obtained. Bias and precision were +1.0 and +2.5 %, respectively, and the limits of agreement ranged from -4.0 to 6.0 %. The finger sensor failed to give a value in four cases, thus providing 136 paired data sets (finger SpO(2) vs. SaO(2)) for analysis. Bias and precision were +1.4 and +4.8 %, respectively, and the limits of agreement ranged from -8.0 to 10.9 %. There were 21 (15 %) outliers for the forehead sensor and 43 (32 %) for the finger sensor (P < 0.001).

Conclusions: Forehead SpO(2) measurements were more accurate than finger SpO(2) when compared with SaO(2) in critically ill patients requiring high-dose vasopressor therapy and should therefore be the preferred method considered.

Citing Articles

Comparing peripheral limb and forehead vital sign monitoring in newborn infants at birth.

Swamy S, Stockwell S, Liu C, Henry C, Shipley L, Ward C Pediatr Res. 2024; .

PMID: 39420152 DOI: 10.1038/s41390-024-03651-0.


Non-invasive assessment of oxygenation status using the oxygen reserve index in dogs.

Zanusso F, De Benedictis G, Zemko P, Bellini L BMC Vet Res. 2023; 19(1):241.

PMID: 37980491 PMC: 10657143. DOI: 10.1186/s12917-023-03804-z.


Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients.

Blanchet M, Mercier G, Delobel A, Nayet E, Bouchard P, Simard S Respir Care. 2023; 68(5):565-574.

PMID: 36596654 PMC: 10171338. DOI: 10.4187/respcare.10582.


Clinical evaluation for the pharyngeal oxygen saturation measurements in shocked patients.

Hassan E, Mohamed S, Hamouda E, Ahmed N BMC Nurs. 2022; 21(1):290.

PMID: 36316710 PMC: 9624017. DOI: 10.1186/s12912-022-01073-z.


Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs.

Nixdorff J, Zablotski Y, Hartmann K, Dorfelt R Front Vet Sci. 2021; 8:643966.

PMID: 33996971 PMC: 8119740. DOI: 10.3389/fvets.2021.643966.


References
1.
Wilson B, Cowan H, Lord J, Zuege D, Zygun D . The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010; 10:9. PMC: 2876142. DOI: 10.1186/1471-227X-10-9. View

2.
Palve H . Reflection and transmission pulse oximetry during compromised peripheral perfusion. J Clin Monit. 1992; 8(1):12-5. DOI: 10.1007/BF01618081. View

3.
Wax D, Rubin P, Neustein S . A comparison of transmittance and reflectance pulse oximetry during vascular surgery. Anesth Analg. 2009; 109(6):1847-9. DOI: 10.1213/ANE.0b013e3181bbc446. View

4.
Smithline H, Rudnitzky N, Macomber S, Blank F . Pulse oximetry using a disposable finger sensor placed on the forehead in hypoxic patients. J Emerg Med. 2009; 39(1):121-5. DOI: 10.1016/j.jemermed.2009.03.019. View

5.
Jubran A . Pulse oximetry. Intensive Care Med. 2004; 30(11):2017-20. DOI: 10.1007/s00134-004-2399-x. View