Measurement of the Microdistribution of Strontium and Lead in Bone Via Benchtop Monochromatic Microbeam X-ray Fluorescence with a Low Power Source

Overview
Date 2012 Jul 17
PMID 22798703
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Determination of the microdistribution of trace elements in bone at low concentrations has previously been performed with proton induced X-ray emission (PIXE), high-energy synchrotron source X-ray fluorescence (XRF) and laser ablation - inductively coupled plasma mass spectrometry (LA-ICP-MS). Several commercial benchtop XRF systems with micrometer-scale resolution are currently available. While providing convenient, non-destructive mapping capability, they appear to lack the sensitivity required for detection of trace elements in biological tissues such as bone. We investigated the application of a prototype benchtop XRF system for the measurement of strontium and lead at physiological levels in bone. Detection of several elements of interest, including Sr and Pb was achieved with an experimental set up based on focused monochromatic microbeam X-ray fluorescence (Mµ-XRF) instrumentation with a low power source (45 W molybdenum tube) coupled to doubly curved crystal (DCC) optics. A cross-section of bone about 5 mm × 8 mm size was mapped with 80-µm resolution showing heterogeneous distribution of Sr and Pb. The data showed that Mµ-XRF coupled to DCC is powerful method for measurement of the spatial distribution of trace elements in bone.

Citing Articles

A novel calibration for L-shell x-ray fluorescence measurements of bone lead concentration using the strontium K/Kratio.

Gherase M, Serna B, Kroeker S Physiol Meas. 2021; 42(4).

PMID: 33857933 PMC: 8177726. DOI: 10.1088/1361-6579/abf886.


Comparison of bone lead measured via portable x-ray fluorescence across and within bones.

Specht A, Dickerson A, Weisskopf M Environ Res. 2019; 172:273-278.

PMID: 30822560 PMC: 6511307. DOI: 10.1016/j.envres.2019.02.031.


Improvements and reproducibility of an optimal grazing-incidence position method to L-shell x-ray fluorescence measurements of lead in bone and soft tissue phantoms.

Gherase M, Al-Hamdani S Biomed Phys Eng Express. 2019; 4(6).

PMID: 30631485 PMC: 6322552. DOI: 10.1088/2057-1976/aae300.


X-ray fluorescence imaging of metals and metalloids in biological systems.

Zhang R, Li L, Sultanbawa Y, Xu Z Am J Nucl Med Mol Imaging. 2018; 8(3):169-188.

PMID: 30042869 PMC: 6056246.


Portable XRF Technology to Quantify Pb in Bone In Vivo.

Specht A, Weisskopf M, Nie L J Biomark. 2015; 2014:398032.

PMID: 26317033 PMC: 4437356. DOI: 10.1155/2014/398032.


References
1.
Boivin G, Deloffre P, Perrat B, Panczer G, Boudeulle M, Mauras Y . Strontium distribution and interactions with bone mineral in monkey iliac bone after strontium salt (S 12911) administration. J Bone Miner Res. 1996; 11(9):1302-11. DOI: 10.1002/jbmr.5650110915. View

2.
Bellis D, Hetter K, Verostek M, Parsons P . Characterization of candidate reference materials for bone lead via interlaboratory study and double isotope dilution mass spectrometry. J Anal At Spectrom. 2012; 23(3):298-308. PMC: 3477175. DOI: 10.1039/B710808G. View

3.
Todd A, Chettle D . In vivo X-ray fluorescence of lead in bone: review and current issues. Environ Health Perspect. 1994; 102(2):172-7. PMC: 1567203. DOI: 10.1289/ehp.94102172. View

4.
Pors Nielsen S . The biological role of strontium. Bone. 2004; 35(3):583-8. DOI: 10.1016/j.bone.2004.04.026. View

5.
Bellis D, Hetter K, Jones J, Amarasiriwardena D, Parsons P . Calibration of laser ablation inductively coupled plasma mass spectrometry for quantitative measurements of lead in bone. J Anal At Spectrom. 2012; 21(9):948-954. PMC: 3402240. DOI: 10.1039/B603435G. View