» Articles » PMID: 22666645

The Reliability and Concurrent Validity of Shoulder Mobility Measurements Using a Digital Inclinometer and Goniometer: a Technical Report

Overview
Date 2012 Jun 6
PMID 22666645
Citations 101
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose/aim: This study investigated the intrarater reliability and concurrent validity of active shoulder mobility measurements using a digital inclinometer and goniometer.

Materials/methods: Two investigators used a goniometer and digital inclinometer to measure shoulder flexion, abduction, internal and external rotation on 30 asymptomatic participants in a blinded repeated measures design.

Results: Excellent intrarater reliability was present with Intraclass Correlation Coefficients (ICC- 3,k) for goniometry ≥ 0.94 and digital inclinometry ≥ 0.95. The concurrent validity between goniometry and digital inclinometry was good with ICC (3,k) values of ≥ 0.85. The 95% limits of agreement suggest that the difference between these two measurement instruments can be expected to range from 2° to 20°.

Conclusions: The results cautiously support the interchangeable use of goniometry and digital inclinometer for measuring shoulder mobility measurements. Although reliable, clinicians should consider the 95% limits of agreement when using these instruments interchangeably as clinically significant differences are likely to be present.

Level Of Evidence: 2b.

Citing Articles

Baseline Characteristics of Participants in the Alberta Cancer Exercise Hybrid Effectiveness-Implementation Study: A Wake-Up Call for Action.

McNeely M, Shallwani S, Williamson T, Sellar C, Gobeil E, Joy A Cancers (Basel). 2025; 17(5).

PMID: 40075620 PMC: 11898868. DOI: 10.3390/cancers17050772.


Exploring the Potential of AI-Assisted Technology in Joint Range-of-Motion Measurements: A Reliability Study.

Lee G, Tan E Medicina (Kaunas). 2025; 61(1).

PMID: 39859101 PMC: 11766890. DOI: 10.3390/medicina61010119.


Reliability of sitting posture between physical therapist video-based evaluation and SMART IMU system using rapid upper limb assessment (RULA).

Tanthuwapathom R, Manupibul U, Jarumethitanont W, Limroongreungrat W, Ongwattanakul S, Charoensuk W Sci Rep. 2025; 15(1):1441.

PMID: 39789041 PMC: 11717903. DOI: 10.1038/s41598-025-85159-z.


Efficacy of Transcutaneous Pulsed Radiofrequency Treatment in Subacromial Impingement Syndrome: A Randomized Controlled Study.

Karasu A, Kilic A, Karaoglan B J Clin Med. 2024; 13(23).

PMID: 39685920 PMC: 11642105. DOI: 10.3390/jcm13237462.


Effects of Lumbar Stabilization on Scapular Muscle Activity, Activation Onset Time, and Kinematics in Individuals with Scapular Dyskinesis.

Sungkue S, Sakulsriprasert P, Vongsirinavarat M, Utsahachant N, Jensen M J Hum Kinet. 2024; 94:23-35.

PMID: 39563765 PMC: 11571471. DOI: 10.5114/jhk/186972.


References
1.
Laupattarakasem W, Sirichativapee W, Kowsuwon W, Sribunditkul S, Suibnugarn C . Axial rotation gravity goniometer. A simple design of instrument and a controlled reliability study. Clin Orthop Relat Res. 1990; (251):271-4. View

2.
Di Fabio R, Boissonnault W . Physical therapy and health-related outcomes for patients with common orthopaedic diagnoses. J Orthop Sports Phys Ther. 1998; 27(3):219-30. DOI: 10.2519/jospt.1998.27.3.219. View

3.
Shrout P, Fleiss J . Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979; 86(2):420-8. DOI: 10.1037//0033-2909.86.2.420. View

4.
Bovens A, van Baak M, Vrencken J, Wijnen J, Verstappen F . Variability and reliability of joint measurements. Am J Sports Med. 1990; 18(1):58-63. DOI: 10.1177/036354659001800110. View

5.
Weir J . Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005; 19(1):231-40. DOI: 10.1519/15184.1. View