Discrepancies Between Capillary Glucose Measurements and Traditional Laboratory Assessments in Both Shock and Non-shock States After Trauma
Overview
Authors
Affiliations
Introduction: The purpose of this study was to analyze the accuracy of capillary blood glucose (CBG) against laboratory blood glucose (LBG) in critically ill trauma patients during the shock state.
Methods: All critically ill trauma patients admitted to the Surgical Intensive Care Unit at the Los Angeles County + University of Southern California Medical Center requiring blood glucose monitoring from January 2007 to December 2008 were included. Accuracy of CBG was compared against LBG during shock and non-shock states. Shock was defined as either systolic blood pressure <90 mm Hg or mean arterial pressure <70 mm Hg and the need for vasopressor therapy. The Bland-Altman method was used to determine the agreement between CBG and LBG during shock and non-shock states. CBG values were considered to disagree significantly with LBG values when the difference exceeded 15%.
Results: During the 2-y study period, a total of 1215 patients were admitted to the Surgical Intensive Care Unit. Overall, the mean age was 38.4 ± 20.9 y, 79.6% (967) were male, and 75.0% (911) sustained blunt trauma. A total of 1935 paired samples of CBG and LBG were included in this analysis (367 during shock and 1568 during non-shock). During shock, the mean difference between CBG and LBG levels was 13.4 mg/dL (95% CI, -15.4 to 42.2 mg/dL), and the limits of agreement were -27.1 and 53.9 mg/dL. A total of 136 CBG values (37.1%) differed from the LBG values by more than 15%. During non-shock, the mean difference between CBG and LBG levels was 12.6 mg/dL (95% CI, -19.9 to 32.5 mg/dL), and the limits of agreement were -20.6 and 45.8 mg/dL. A total of 639 CGB values (40.8%) differed from the LBG values by more than 15%. Agreement was lowest among hypoglycemic readings in both shock and non-shock states.
Conclusion: There is poor correlation between the capillary and laboratory glucose values in both shock and non-shock states.
Reliability and criterion validity of self-measured waist, hip, and neck circumferences.
Barrios P, Martin-Biggers J, Quick V, Byrd-Bredbenner C BMC Med Res Methodol. 2016; 16:49.
PMID: 27145829 PMC: 4855335. DOI: 10.1186/s12874-016-0150-2.
Munekage M, Yatabe T, Sakaguchi M, Kitagawa H, Tamura T, Namikawa T J Artif Organs. 2016; 19(2):159-66.
PMID: 26721825 DOI: 10.1007/s10047-015-0877-2.
Yaraghi A, Mood N, Dolatabadi L Adv Biomed Res. 2015; 4:247.
PMID: 26693472 PMC: 4685638. DOI: 10.4103/2277-9175.170242.
Suitability of capillary blood glucose analysis in patients receiving vasopressors.
Ellis M, Benjamin K, Cornell M, Decker K, Farrell D, McGugan L Am J Crit Care. 2013; 22(5):423-9.
PMID: 23996422 PMC: 4324832. DOI: 10.4037/ajcc2013692.