Oscillometric Estimation of Central Blood Pressure: Validation of the Mobil-O-Graph in Comparison with the SphygmoCor Device
Overview
Authors
Affiliations
Background: Hypertension is a major risk factor for a wide range of cardiovascular diseases and is typically identified by measuring blood pressure (BP) at the brachial artery. Although such a measurement may accurately determine diastolic BP, systolic BP is not reflected accurately. Current noninvasive techniques for assessing central aortic BP require additional recording of an arterial pressure wave using a high-fidelity applanation tonometer. Within one measurement cycle, the Mobil-O-Graph BP device uses brachial oscillometric BP waves for a noninvasive estimation of central BP. We therefore validated the Mobil-O-Graph against the SphygmoCor device, which is widely known as the commonly used approach for a noninvasive estimation of central BP.
Methods: For each individual, we compared three readings of the central BP values obtained by the Mobil-O-Graph and SphygmoCor device consecutively. One hundred individuals (mean age 56.1 ± 15.4 years) were recruited for measurement.Differences between the central BP values of the test device and the SphygmoCor device were calculated for each measurement.
Results: The mean difference (95% confidence interval) for the estimated central systolic BP between both devices was -0.6 ± 3.7 mmHg. Comparison of the central BP values measured by the two devices showed a statistically significant linear correlation (R=0.91, P<0.0001). The mean between-method difference was 0.50 mmHg for central systolic BP estimation. The intrarater reproducibility between both the devices was also comparable. Bland and Altman analyses showed that the mean differences (95% confidence interval) between repeated measurements were 1.89 (0.42-3.36) mmHg and 1.36 (-0.16 to 2.83) mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively. Thus, neither of these differences was statistically significantly different from 0. The limits of agreement were -16.34 to 19.73 and -15.23 to 17.17 mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively.
Conclusion: Oscillometric noninvasive estimation of central BP with the Mobil-O-Graph BP device is as effective as using the well-established SphygmoCor applanation tonometry device. In comparison, the Mobil-O-Graph combines the widespread benefits of brachial BP measurement and also provides central BP within one measurement.
Tamborini A, Aghilinejad A, Gharib M Proc Natl Acad Sci U S A. 2025; 122(9):e2416006122.
PMID: 40009644 PMC: 11892652. DOI: 10.1073/pnas.2416006122.
Sgambat K, Amatya K, Vogel S, Clauss S, Moudgil A Pediatr Nephrol. 2025; .
PMID: 39808333 DOI: 10.1007/s00467-024-06635-8.
Dalton-Alves F, Araujo M, Lucena B, Souto G, Lopes D, Lucena M BMJ Open. 2025; 14(12):e084736.
PMID: 39806596 PMC: 11664350. DOI: 10.1136/bmjopen-2024-084736.
Haferanke J, Baumgartner L, Willinger L, Schulz T, Muhlbauer F, Engl T CJC Open. 2024; 6(12):1549-1557.
PMID: 39735949 PMC: 11681355. DOI: 10.1016/j.cjco.2024.09.005.
Thaxton C, Kano M, Mendes-Pinto D, Navarro T, Nishibe T, Dardik A JVS Vasc Sci. 2024; 5:100209.
PMID: 39677517 PMC: 11639741. DOI: 10.1016/j.jvssci.2024.100209.