Flow Diverters Failing to Occlude Experimental Bifurcation or Curved Sidewall Aneurysms: an in Vivo Study in Canines

Overview
Journal J Neurosurg
Specialty Neurosurgery
Date 2012 May 8
PMID 22559845
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Object: Flow diverters (FDs) are increasingly used to treat complex intracranial aneurysms, but preclinical studies that could guide clinical applications are lacking. The authors designed a modular aneurysm model in canines to address this problem.

Methods: Three variants of one modular aneurysm model were constructed in 21 animals. Sidewall (n=5), curved sidewall (n=5), and end-wall bifurcation (n=7) aneurysms were treated with prototype 36-wire FDs. Four more end-wall bifurcation aneurysms were treated with prototype 48-wire lower-porosity FDs. Angiographic results postimplantation and at 3 months were scored with an ordinal scale. Animals were euthanized at 3 (n=17) or 6 (n=3) months, and the FD covering the aneurysm ostium was photographed to analyze metallic porosity and amount of neointima formation.

Results: Straight sidewall aneurysms were better occluded than curved sidewall and end-wall bifurcation aneurysms at the 3-month angiography follow-up (p=0.010). Flow diverters failed to occlude curved sidewall aneurysms (n=0/5) and all but one (n=1/7) end-wall bifurcation aneurysm. Angiographic results were no better (n=0/4) using a 48-wire FD (p=0.788). Branches jailed by the FD (n=16) remained patent in all cases. Metallic porosity was decreased (p=0.014) and neointimal closure of the aneurysm ostium was more complete (p=0.040) in sidewall aneurysms than in curved or bifurcation variants of the model.

Conclusions: Flow diverters may succeed in treating straight sidewall aneurysms, but the same device repeatedly fails to occlude curved sidewall and end-wall bifurcation aneurysms. In vivo studies can be designed to test basic principles that, once validated, may serve to guide clinical use of new devices.

Citing Articles

Initial Experience with the New DERIVO Mini Embolisation Device for the Treatment of Intracranial Aneurysms.

Balci S, Cay F, Uysal A, Arat A Brain Sci. 2024; 14(9).

PMID: 39335406 PMC: 11430613. DOI: 10.3390/brainsci14090911.


A Case of Horizontal Stent-assisted Coiling for an Aneurysm Arising from Fenestration of the Vertebral Artery: A Technical Case Report.

Miyata H, Ninomiya F, Fujita T, Nonoyama Y, Nakazawa T, Fukao S Neurol Med Chir (Tokyo). 2023; 63(6):258-263.

PMID: 37005245 PMC: 10325777. DOI: 10.2176/jns-nmc.2022-0309.


Imaging of intracranial aneurysms in animals: a systematic review of modalities.

Cayron A, Morel S, Allemann E, Bijlenga P, Kwak B Neurosurg Rev. 2023; 46(1):56.

PMID: 36786880 PMC: 9928939. DOI: 10.1007/s10143-023-01953-1.


Structural changes in the side branches and the circle of Willis following the use of flowdiverting stents.

Onan H, Pehlivan U, Sozutok S, Kose S, Akgul E Turk J Med Sci. 2022; 52(4):965-974.

PMID: 36326385 PMC: 10387835. DOI: 10.55730/1300-0144.5397.


Small Unruptured Intracranial Aneurysms Can Be Effectively Treated With Flow-Diverting Devices.

Li L, Gao B, Shao Q, Zhang G, Wang Z, Li T Front Neurol. 2022; 13:913653.

PMID: 35707031 PMC: 9190436. DOI: 10.3389/fneur.2022.913653.