Purpose: The aim of this study was to compare low-dose imaging with gadobutrol and gadoterate meglumine (Gd-DOTA) for evaluation of renal artery stenosis with 3-T magnetic resonance angiography (MRA) in a swine model.

Method And Materials: A total of 12 experimental animals were evaluated using equivalently dosed gadobutrol and Gd-DOTA for time-resolved and static imaging. For time-resolved imaging, the time-resolved imaging with stochastic trajectories (TWIST) technique (temporal footprint, 4.4 seconds) was used; a dose of 1 mL of gadobutrol was injected at 2 mL/s and a dose of 2 mL of Gd-DOTA was injected at both 2 and 4 mL/s. For a separate static acquisition, doses were doubled. The static scans were used for stenosis gradation and the time-resolved scans for comparison of enhancement dynamics, signal-to-noise ratio (SNR), and qualitative assessments.

Results: The average magnitude of difference in the stenosis measurements with static gadobutrol scans relative to digital subtraction intra-arterial catheter angiography (mean [SD], 7.4% [5.6%]) was less than with both the 2 mL/s (10.6% [6.2%]) and 4 mL/s (11.5% [7.8%]) Gd-DOTA MRA protocols. On time-resolved scans, peak signal-to-noise ratio was greatest with the gadobutrol protocol (P < 0.05), and the gadobutrol TWIST scan was preferred to the TWIST Gd-DOTA scan in terms of image quality and stenosis visualization in every case for every reader.

Conclusion: Low-dose gadobutrol (~0.05 mmoL/kg) contrast-enhanced MRA results in improved accuracy of renal artery stenosis assessments relative to equivalently dosed Gd-DOTA at 3 T.

Citing Articles

High spatial resolution time-resolved magnetic resonance angiography of lower extremity tumors at 3T: Comparison with computed tomography angiography.

Wu G, Jin T, Li T, Morelli J, Li X Medicine (Baltimore). 2016; 95(37):e4894.

PMID: 27631262 PMC: 5402605. DOI: 10.1097/MD.0000000000004894.


The diagnostic value of time-resolved MR angiography with Gadobutrol at 3 T for preoperative evaluation of lower extremity tumors: Comparison with computed tomography angiography.

Wu G, Jin T, Li T, Li X Eur J Radiol Open. 2016; 3:239-44.

PMID: 27622202 PMC: 5009234. DOI: 10.1016/j.ejro.2016.08.006.