» Articles » PMID: 22536333

The OSCAR-IB Consensus Criteria for Retinal OCT Quality Assessment

Overview
Journal PLoS One
Date 2012 Apr 27
PMID 22536333
Citations 275
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Retinal optical coherence tomography (OCT) is an imaging biomarker for neurodegeneration in multiple sclerosis (MS). In order to become validated as an outcome measure in multicenter studies, reliable quality control (QC) criteria with high inter-rater agreement are required.

Methods/principal Findings: A prospective multicentre study on developing consensus QC criteria for retinal OCT in MS: (1) a literature review on OCT QC criteria; (2) application of these QC criteria to a training set of 101 retinal OCT scans from patients with MS; (3) kappa statistics for inter-rater agreement; (4) identification reasons for inter-rater disagreement; (5) development of new consensus QC criteria; (6) testing of the new QC criteria on the training set and (7) prospective validation on a new set of 159 OCT scans from patients with MS. The inter-rater agreement for acceptable scans among OCT readers (n = 3) was moderate (kappa 0·45) based on the non-validated QC criteria which were entirely based on the ophthalmological literature. A new set of QC criteria was developed based on recognition of: (O) obvious problems, (S) poor signal strength, (C) centration of scan, (A) algorithm failure, (R) retinal pathology other than MS related, (I) illumination and (B) beam placement. Adhering to these OSCAR-IB QC criteria increased the inter-rater agreement to kappa from moderate to substantial (0.61 training set and 0.61 prospective validation).

Conclusions: This study presents the first validated consensus QC criteria for retinal OCT reading in MS. The high inter-rater agreement suggests the OSCAR-IB QC criteria to be considered in the context of multicentre studies and trials in MS.

Citing Articles

A Novel Nomogram Integrating Retinal Microvasculature and Clinical Indicators for Individualized Prediction of Early Neurological Deterioration in Single Subcortical Infarction.

Ye C, Kwapong W, Cao L, Xu H, Wang Y, Yan Y CNS Neurosci Ther. 2025; 31(3):e70337.

PMID: 40074710 PMC: 11903217. DOI: 10.1111/cns.70337.


Longitudinal analysis of retinal and choriocapillaris in patients with multiple sclerosis: a 3-year study.

Bostan M, Li C, Cheong J, Wong D, Tan B, Ismail M J Neurol. 2025; 272(3):237.

PMID: 40025301 PMC: 11872746. DOI: 10.1007/s00415-025-12930-7.


Lower Retinal Arteriolar Density Is Associated With Higher Cerebral Small Vessel Disease Burden: An Optical Coherence Tomography Angiography Study.

Cao L, Wang H, Hao J, Kwapong W, Xiong Z, Wang R Brain Behav. 2025; 15(2):e70342.

PMID: 39957423 PMC: 11830995. DOI: 10.1002/brb3.70342.


OCT-Based Retina Assessment Reflects Visual Impairment in Multiple Sclerosis.

Stolowy N, Gutmann L, Lupke M, David T, Dorr M, Mayer C Invest Ophthalmol Vis Sci. 2025; 66(2):39.

PMID: 39946137 PMC: 11827617. DOI: 10.1167/iovs.66.2.39.


New Method of Early RRMS Diagnosis Using OCT-Assessed Structural Retinal Data and Explainable Artificial Intelligence.

Ortiz M, Pueyo A, Dongil F, Boquete L, Sanchez-Morla E, Barea R Transl Vis Sci Technol. 2025; 14(2):14.

PMID: 39928305 PMC: 11812612. DOI: 10.1167/tvst.14.2.14.


References
1.
Balk L, de Vries-Knoppert W, Petzold A . A simple sign for recognizing off-axis OCT measurement beam placement in the context of multicentre studies. PLoS One. 2012; 7(11):e48222. PMC: 3493550. DOI: 10.1371/journal.pone.0048222. View

2.
Huang J, Liu X, Wu Z, Sadda S . Image quality affects macular and retinal nerve fiber layer thickness measurements on fourier-domain optical coherence tomography. Ophthalmic Surg Lasers Imaging. 2011; 42(3):216-21. DOI: 10.3928/15428877-20110324-01. View

3.
Barkhof F, Calabresi P, Miller D, Reingold S . Imaging outcomes for neuroprotection and repair in multiple sclerosis trials. Nat Rev Neurol. 2009; 5(5):256-66. DOI: 10.1038/nrneurol.2009.41. View

4.
Vizzeri G, Bowd C, Medeiros F, Weinreb R, Zangwill L . Effect of signal strength and improper alignment on the variability of stratus optical coherence tomography retinal nerve fiber layer thickness measurements. Am J Ophthalmol. 2009; 148(2):249-255.e1. PMC: 2848173. DOI: 10.1016/j.ajo.2009.03.002. View

5.
Domalpally A, Danis R, Zhang B, Myers D, Kruse C . Quality issues in interpretation of optical coherence tomograms in macular diseases. Retina. 2009; 29(6):775-81. DOI: 10.1097/IAE.0b013e3181a0848b. View