» Articles » PMID: 22535813

Acutrak Versus Herbert Screw Fixation for Scaphoid Non-union and Delayed Union

Overview
Publisher Sage Publications
Specialty Orthopedics
Date 2012 Apr 27
PMID 22535813
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare the treatment outcome of Acutrak versus Herbert screw fixation for scaphoid non-union and delayed union.

Methods: Records of 132 patients who underwent Herbert screw fixation (n=61) or Acutrak screw fixation (n=71) with or without bone grafting for scaphoid non-union and delayed union by a single surgeon were reviewed. The most common fracture site was the waist of the scaphoid (n=95), followed by the proximal pole (n=31) and the distal pole (n=6). Screw placement was considered accurate (n=120) when the screw was placed in the central one-third (axially) of the scaphoid; otherwise it was eccentric (n=12). Bone union was assessed radiographically and clinically. Functional outcome was assessed using the modified Mayo wrist score.

Results: Respectively in the Herbert and Acutrak screw groups, the mean patient ages were 25.3 and 27.3 years (p=0.28), the mean intervals between injury and screw fixation were 12.2 and 17 months (p=0.38), the mean durations to bone union were 2.1 and 1.8 months (p=0.63), and the union rates were 77% and 93% (p=0.01). The union rate was significantly higher in fractures of the waist of the scaphoid than in the proximal and distal poles (94% vs. 71% vs. 33%, p=0.001). The union rate was significantly higher when the screw was placed accurately (axially) than eccentrically (Herbert screw: 84% vs. 40%, p=0.006; Acutrak screw: 96% vs. 0%, p=0.004). 84% of the Herbert screws were placed axially, compared to 97% for the Acutrak screws. Respectively, 67% and 85% of patients had satisfactory functional outcomes (p=0.03), whereas 23% and 7% of the patients had persistent non-union (p=0.05).

Conclusion: The Acutrak screw enabled more accurate screw placement and achieved higher union rates and modified Mayo wrist scores than the Herbert screw did.

Citing Articles

Decreased Complication Rate Using Partially Threaded Compared With Fully Threaded Compression Screws in 4 Corner Arthrodesis: A Retrospective Study.

Hanna T, Smith N, Akalonu C, Simmons J, Nehete P, Dodson M Hand (N Y). 2023; 19(8):1214-1222.

PMID: 37128854 PMC: 11536756. DOI: 10.1177/15589447231167883.


Volar Plating of Scaphoid Fractures: A Retrospective Case Series.

Lemke K, Mannambeth R, Carman C, Csongvay S Hand (N Y). 2022; 18(2_suppl):46S-51S.

PMID: 35227110 PMC: 10052624. DOI: 10.1177/15589447221075674.


How to ease the fixation of carpal bones? Twenty-four lesions treated by 1,5 mm headless compression screws.

Poggetti A, Suardi C, Lauri G, Espen D, Bigazzi P, Pfanner S J Clin Orthop Trauma. 2021; 18:56-60.

PMID: 33996449 PMC: 8099553. DOI: 10.1016/j.jcot.2021.03.022.


Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws.

Filli L, Marcon M, Scholz B, Calcagni M, Finkenstadt T, Andreisek G Skeletal Radiol. 2014; 43(12):1705-12.

PMID: 25164608 DOI: 10.1007/s00256-014-1986-3.


Insertion profiles of 4 headless compression screws.

Hart A, Harvey E, Lefebvre L, Barthelat F, Rabiei R, Martineau P J Hand Surg Am. 2013; 38(9):1728-34.