Commercially Pure Porous Titanium (CPPTi) can be used for surgical implants to avoid the stress shielding effect due to the mismatch between the mechanical properties of titanium and bone. Most researchers in this area deal with randomly distributed pores or simple architectures in titanium alloys. The control of porosity, pore size and distribution is necessary to obtain implants with mechanical properties close to those of bone and to ensure their osseointegration. The aim of the present work was therefore to develop and characterize such a specific porous structure. First of all, the properties of titanium made by Selective Laser Melting (SLM) were characterized through experimental testing on bulk specimens. An elementary pattern of the porous structure was then designed to mimic the orthotropic properties of the human bone following several mechanical and geometrical criteria. Finite Element Analysis (FEA) was used to optimize the pattern. A porosity of 53% and pore sizes in the range of 860 to 1500 μm were finally adopted. Tensile tests on porous samples were then carried out to validate the properties obtained numerically and identify the failure modes of the samples. Finally, FE elastoplastic analyses were performed on the porous samples in order to propose a failure criterion for the design of porous substitutes.

Citing Articles

Surgical Planning and 3D-Printed Mesh Implant for Effective Bone Gap Management: A Case Report.

Kachare A, Goregaonkar A, Purohit S, Munde K, Renthlei L, Gaur B J Orthop Case Rep. 2024; 14(11):203-207.

PMID: 39524287 PMC: 11546020. DOI: 10.13107/jocr.2024.v14.i11.4968.


Anisotropy of Additively Manufactured Metallic Materials.

Huangfu B, Liu Y, Liu X, Wu X, Bai H Materials (Basel). 2024; 17(15).

PMID: 39124317 PMC: 11312974. DOI: 10.3390/ma17153653.


Biomechanical analysis of titanium-alloy and biodegradable implants in dual plate osteosynthesis for AO/ASIF type 33-C2 fractures.

Hu M, Li M, Ma R, Li X, Ren X, Du L Heliyon. 2024; 10(4):e26213.

PMID: 38404819 PMC: 10884484. DOI: 10.1016/j.heliyon.2024.e26213.


The effect of porous compliance bushings in a dental implant on the distribution of occlusal loads.

Mlynarek-Zak K, Zmudzki J Sci Rep. 2024; 14(1):1607.

PMID: 38238380 PMC: 10796672. DOI: 10.1038/s41598-024-51429-5.


Development of Bioactive Scaffolds for Orthopedic Applications by Designing Additively Manufactured Titanium Porous Structures: A Critical Review.

Kiselevskiy M, Anisimova N, Kapustin A, Ryzhkin A, Kuznetsova D, Polyakova V Biomimetics (Basel). 2023; 8(7).