Overcoming the Effects of Matrix Interference in the Measurement of Urine Protein Analytes

Overview
Publisher Sage Publications
Date 2012 Mar 10
PMID 22403482
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Using multiplex bead assays to measure urine proteins has a great potential for biomarker discovery, but substances in urine (the matrix) can interfere with assay measurements. By comparing the recovery of urine spiked with known quantities of several common analytes, this study demonstrated that the urine matrix variably interfered with the accurate measurement of low abundance proteins. Dilution of the urine permitted a more accurate measure of these proteins, equivalent to the standard dilution technique when the diluted analytes were above the limits of detection of the assay. Therefore, dilution can be used as an effective technique for over-coming urine matrix effects in urine immunoassays. These results may be applicable to other biological fluids in which matrix components interfere with assay performance.

Citing Articles

Lithographically defined encoded magnetic heterostructures for the targeted screening of kidney cancer.

Leulmi Pichot S, Vemulkar T, Verheyen J, Wallis L, Jones J, Stewart A Nanoscale Adv. 2023; 6(1):276-286.

PMID: 38125591 PMC: 10729922. DOI: 10.1039/d3na00701d.


Detection and Quantification of the Capsular Polysaccharide of Burkholderia pseudomallei in Serum and Urine Samples from Melioidosis Patients.

DeMers H, Nualnoi T, Thorkildson P, Hau D, Hannah E, Green H Microbiol Spectr. 2022; 10(4):e0076522.

PMID: 35924843 PMC: 9430648. DOI: 10.1128/spectrum.00765-22.


Development and Validation of a Sensitive and Robust Multiplex Antigen Capture Assay to Quantify Streptococcus pneumoniae Serotype-Specific Capsular Polysaccharides in Urine.

Rajam G, Zhang Y, Antonello J, Grant-Klein R, Cook L, Panemangalore R mSphere. 2022; 7(4):e0011422.

PMID: 35913133 PMC: 9429912. DOI: 10.1128/msphere.00114-22.


Urinary immunoglobulins in viral diagnosis: An overview.

Mohandas S, Balan S, Mourya D Indian J Med Res. 2022; 155(1):11-21.

PMID: 35859424 PMC: 9552372. DOI: 10.4103/ijmr.IJMR_808_18.


A Novel Propidium Monoazide-Based PCR Assay Can Measure Viable Uropathogenic and .

Lee A, Lamanna O, Ishida K, Hill E, Nguyen A, Hsieh M Front Cell Infect Microbiol. 2022; 12:794323.

PMID: 35178354 PMC: 8844370. DOI: 10.3389/fcimb.2022.794323.


References
1.
Zhou H, Hewitt S, Yuen P, Star R . Acute Kidney Injury Biomarkers - Needs, Present Status, and Future Promise. Nephrol Self Assess Program. 2008; 5(2):63-71. PMC: 2603572. View

2.
Basilicata P, Miraglia N, Pieri M, Acampora A, Soleo L, Sannolo N . Application of the standard addition approach for the quantification of urinary benzene. J Chromatogr B Analyt Technol Biomed Life Sci. 2005; 818(2):293-9. DOI: 10.1016/j.jchromb.2005.01.013. View

3.
Sviridov D, Hortin G . Urine albumin measurement: effects of urine matrix constituents. Clin Chim Acta. 2009; 404(2):140-3. DOI: 10.1016/j.cca.2009.03.034. View

4.
Wood M, Nordone S, Vaden S, Breitschwerdt E . Assessment of urine solute and matrix effects on the performance of an enzyme-linked immunosorbent assay for measurement of interleukin-6 in dog urine. J Vet Diagn Invest. 2011; 23(2):316-20. PMC: 4142517. DOI: 10.1177/104063871102300219. View

5.
Fichorova R, Richardson-Harman N, Alfano M, Belec L, Carbonneil C, Chen S . Biological and technical variables affecting immunoassay recovery of cytokines from human serum and simulated vaginal fluid: a multicenter study. Anal Chem. 2008; 80(12):4741-51. PMC: 2646866. DOI: 10.1021/ac702628q. View