Immediate Effects of a New Microprocessor-controlled Prosthetic Knee Joint: a Comparative Biomechanical Evaluation
Overview
Affiliations
Objective: To investigate the immediate biomechanical effects after transition to a new microprocessor-controlled prosthetic knee joint.
Design: Intervention cross-over study with repeated measures. Only prosthetic knee joints were changed.
Setting: Motion analysis laboratory.
Participants: Men (N=11; mean age ± SD, 36.7±10.2y; Medicare functional classification level, 3-4) with unilateral transfemoral amputation.
Interventions: Two microprocessor-controlled prosthetic knee joints: C-Leg and a new prosthetic knee joint, Genium.
Main Outcome Measures: Static prosthetic alignment, time-distance parameters, kinematic and kinetic parameters, and center of pressure.
Results: After a half-day training and an additional half-day accommodation, improved biomechanical outcomes were demonstrated by the Genium: lower ground reaction forces at weight acceptance during level walking at various velocities, increased swing phase flexion angles during walking on a ramp, and level walking with small steps. Maximum knee flexion angle during swing phase at various velocities was nearly equal for Genium. Step-over-step stair ascent with the Genium knee was more physiologic as demonstrated by a more equal load distribution between the prosthetic and contralateral sides and a more natural gait pattern. When descending stairs and ramps, knee flexion moments with the Genium tended to increase. During quiet stance on a decline, subjects using Genium accepted higher loading of the prosthetic side knee joint, thus reducing same side hip joint loading as well as postural sway.
Conclusions: In comparision to the C-Leg, the Genium demonstrated immediate biomechanical advantages during various daily ambulatory activities, which may lead to an increase in range and diversity of activity of people with above-knee amputations. Results showed that use of the Genium facilitated more natural gait biomechanics and load distribution throughout the affected and sound musculoskeletal structure. This was observed during quiet stance on a decline, walking on level ground, and walking up and down ramps and stairs.
Cheng S, Laubscher C, Gregg R IEEE Trans Neural Syst Rehabil Eng. 2024; 32:3891-3901.
PMID: 39446547 PMC: 11608573. DOI: 10.1109/TNSRE.2024.3485643.
Tanaka Y, Ueno T Cureus. 2023; 15(10):e46566.
PMID: 37936990 PMC: 10626260. DOI: 10.7759/cureus.46566.
A Scientometric Analysis and Visualization of Prosthetic Foot Research Work: 2000 to 2022.
Shi Q, Yick K, Wu J, Huang X, Tse C, Chan M Bioengineering (Basel). 2023; 10(10).
PMID: 37892868 PMC: 10604169. DOI: 10.3390/bioengineering10101138.
Geil M, Safaeepour Z Front Rehabil Sci. 2023; 4:1227870.
PMID: 37691910 PMC: 10485386. DOI: 10.3389/fresc.2023.1227870.
Benefits of a microprocessor-controlled prosthetic foot for ascending and descending slopes.
Ernst M, Altenburg B, Schmalz T, Kannenberg A, Bellmann M J Neuroeng Rehabil. 2022; 19(1):9.
PMID: 35090505 PMC: 8796559. DOI: 10.1186/s12984-022-00983-y.