Proof-of-concept studies that display the potential of using a glucose-sensitive hydrogel as a continuous glucose sensor are presented. The swelling ratio, porosity, and diffusivity of the hydrogel increased with glucose concentration. In glucose solutions of 50, 100, 200, and 300 mg/dL, the hydrogel swelling ratios were 4.9, 12.3, 15.9, and 21.7, respectively, and the swelling was reversible. The impedance across the hydrogel depended solely on the thickness and had an average increase of 47 Ω/mm. The hydrogels exposed to a hyperglycemic solution were more porous than the hydrogels exposed to a normal glycemic solution. The diffusivity of 390 Da MW fluorescein isothiocyanate in hydrogels exposed to normal and hyperglycemic solutions was examined using fluorescence recovery after photobleaching and was found to be 9.3 × 10(-14) and 41.4 × 10(-14) m(2)/s, respectively, compared to 6.2 × 10(-10) m(2)/s in glucose solution. There was no significant difference between the permeability of hydrogels in normal and hyperglycemic glucose solutions with averages being 5.26 × 10(-17) m(2) and 5.80 × 10(-17) m(2), respectively, which resembles 2-4% agarose gels. A prototype design is presented for continuous intravascular glucose monitoring by attaching a glucose sensor to an FDA-approved stent.

Citing Articles

Advances in the Design of Phenylboronic Acid-Based Glucose-Sensitive Hydrogels.

Morariu S Polymers (Basel). 2023; 15(3).

PMID: 36771883 PMC: 9919422. DOI: 10.3390/polym15030582.


Effects of a synthetic bioactive peptide on neurite growth and nerve growth factor release in chondroitin sulfate hydrogels.

Conovaloff A, Beier B, Irazoqui P, Panitch A Biomatter. 2013; 1(2):165-73.

PMID: 23507745 PMC: 3549887. DOI: 10.4161/biom.17849.

References
1.
Nauman E, Fong K, Keaveny T . Dependence of intertrabecular permeability on flow direction and anatomic site. Ann Biomed Eng. 1999; 27(4):517-24. DOI: 10.1114/1.195. View

2.
Wild S, Roglic G, Green A, Sicree R, King H . Global prevalence of diabetes: estimates for the year 2000 and projections for 2030. Diabetes Care. 2004; 27(5):1047-53. DOI: 10.2337/diacare.27.5.1047. View

3.
. Economic costs of diabetes in the U.S. In 2007. Diabetes Care. 2008; 31(3):596-615. DOI: 10.2337/dc08-9017. View

4.
Chow E, Beier B, Francino A, Chappell W, Irazoqui P . Toward an implantable wireless cardiac monitoring platform integrated with an FDA-approved cardiovascular stent. J Interv Cardiol. 2009; 22(5):479-87. DOI: 10.1111/j.1540-8183.2009.00483.x. View

5.
. Lifetime benefits and costs of intensive therapy as practiced in the diabetes control and complications trial. The Diabetes Control and Complications Trial Research Group. JAMA. 1996; 276(17):1409-15. View