Irradiation of the Prostate and Pelvic Lymph Nodes with an Adaptive Algorithm

Overview
Journal Med Phys
Specialty Biophysics
Date 2012 Feb 11
PMID 22320822
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The simultaneous treatment of pelvic lymph nodes and the prostate in radiotherapy for prostate cancer is complicated by the independent motion of these two target volumes. In this work, the authors study a method to adapt intensity modulated radiation therapy (IMRT) treatment plans so as to compensate for this motion by adaptively morphing the multileaf collimator apertures and adjusting the segment weights.

Methods: The study used CT images, tumor volumes, and normal tissue contours from patients treated in our institution. An IMRT treatment plan was then created using direct aperture optimization to deliver 45 Gy to the pelvic lymph nodes and 50 Gy to the prostate and seminal vesicles. The prostate target volume was then shifted in either the anterior-posterior direction or in the superior-inferior direction. The treatment plan was adapted by adjusting the aperture shapes with or without re-optimizing the segment weighting. The dose to the target volumes was then determined for the adapted plan.

Results: Without compensation for prostate motion, 1 cm shifts of the prostate resulted in an average decrease of 14% in D-95%. If the isocenter is simply shifted to match the prostate motion, the prostate receives the correct dose but the pelvic lymph nodes are underdosed by 14% ± 6%. The use of adaptive morphing (with or without segment weight optimization) reduces the average change in D-95% to less than 5% for both the pelvic lymph nodes and the prostate.

Conclusions: Adaptive morphing with and without segment weight optimization can be used to compensate for the independent motion of the prostate and lymph nodes when combined with daily imaging or other methods to track the prostate motion. This method allows the delivery of the correct dose to both the prostate and lymph nodes with only small changes to the dose delivered to the target volumes.

Citing Articles

Six-degrees-of-freedom pelvic bone monitoring on 2D kV intrafraction images to enable multi-target tracking for locally advanced prostate cancer.

Hewson E, Dillon O, Poulsen P, Booth J, Keall P Med Phys. 2024; 52(1):77-87.

PMID: 39441205 PMC: 11700008. DOI: 10.1002/mp.17465.


Evaluation of knowledge-based planning models for male pelvic CBCT-based online adaptive radiotherapy on conventional linear accelerators.

Tegtmeier R, Clouser E, Laughlin B, Santos Toesca D, Flakus M, Bashir S J Appl Clin Med Phys. 2024; 25(9):e14464.

PMID: 39031902 PMC: 11492302. DOI: 10.1002/acm2.14464.


Varian ethos online adaptive radiotherapy for prostate cancer: Early results of contouring accuracy, treatment plan quality, and treatment time.

Byrne M, Archibald-Heeren B, Hu Y, Teh A, Beserminji R, Cai E J Appl Clin Med Phys. 2021; 23(1):e13479.

PMID: 34846098 PMC: 8803282. DOI: 10.1002/acm2.13479.


Impact of prostate focused alignment on planned pelvic lymph node dose.

Kilian-Meneghin J, Ma T, Kumaraswamy L J Appl Clin Med Phys. 2021; 22(7):27-35.

PMID: 34231945 PMC: 8292696. DOI: 10.1002/acm2.13092.