» Articles » PMID: 22255234

Probabilistic Modeling of Selective Stimulation of the Human Sciatic Nerve with a Flat Interface Nerve Electrode

Overview
Date 2012 Jan 19
PMID 22255234
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Proper ankle control is critical to both standing balance and efficient walking. This study hypothesized that a Flat Interface Nerve Electrode (FINE) placed around the sciatic nerve with a fixed number of contacts at predetermined locations and without a priori knowledge of the nerve's underlying neuroanatomy can selectively control each ankle motion. Models of the human sciatic nerve surrounded by a FINE of varying size were created and used to calculate the probability of selective activation of axons within any arbitrarily designated group of fascicles. Simulations suggest that currently available implantable technology cannot selectively recruit each target plantar flexor individually but can restore plantar flexion or dorsiflexion from a site on the sciatic nerve without spillover to antagonists. Successful activation of individual ankle muscles in 90% of the population can be achieved by utilizing bipolar stimulation and/or by increasing the number of contacts within the cuff.

Citing Articles

Selective Neural Electrical Stimulation of an Injured Facial Nerve Using Chronically Implanted Dual Cuff Electrodes.

Abiri A, Chau S, James N, Goshtasbi K, Birkenbeuel J, Sahyouni R Brain Sci. 2022; 12(11).

PMID: 36358383 PMC: 9688741. DOI: 10.3390/brainsci12111457.

References
1.
Delp S, Anderson F, Arnold A, Loan P, Habib A, John C . OpenSim: open-source software to create and analyze dynamic simulations of movement. IEEE Trans Biomed Eng. 2007; 54(11):1940-50. DOI: 10.1109/TBME.2007.901024. View

2.
Richardson A, McIntyre C, Grill W . Modelling the effects of electric fields on nerve fibres: influence of the myelin sheath. Med Biol Eng Comput. 2000; 38(4):438-46. DOI: 10.1007/BF02345014. View

3.
Tarler M, Mortimer J . Comparison of joint torque evoked with monopolar and tripolar-cuff electrodes. IEEE Trans Neural Syst Rehabil Eng. 2003; 11(3):227-35. DOI: 10.1109/TNSRE.2003.816867. View

4.
. Prevalence of stroke--United States, 2005. MMWR Morb Mortal Wkly Rep. 2007; 56(19):469-74. View

5.
Tyler D, Durand D . Functionally selective peripheral nerve stimulation with a flat interface nerve electrode. IEEE Trans Neural Syst Rehabil Eng. 2003; 10(4):294-303. DOI: 10.1109/TNSRE.2002.806840. View