Validation of Four Automatic Devices for Self-measurement of Blood Pressure According to the International Protocol of the European Society of Hypertension

Overview
Publisher Dove Medical Press
Date 2011 Dec 17
PMID 22174581
Citations 41
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Four oscillometric devices for self-measurement of blood pressure (SBPM) were evaluated according to the European Society of Hypertension (ESH) international protocol and its 2010 revision in four separate studies. The Omron® M2, Omron M3, and Omron M6 measure blood pressure (BP) at the brachial level, while the Omron R2 measures BP at the wrist level.

Methods: The international protocol requires a total number of 33 subjects in which the validation is performed. The Omron M2 and Omron R2 were validated in 2009 according to the ESH international protocol, while the Omron M3 and Omron M6 were validated in 2010-2011 according to the 2010 ESH international protocol revision. The protocol procedures were followed precisely.

Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were 2.7 ± 5.0 and -1.4 ± 3.2 mmHg for systolic and diastolic BP, respectively, using the Omron M2 device, and 1.7 ± 3.2 and -0.9 ± 2.6 mmHg using the Omron M3, 1.6 ± 2.9 and -0.9 ± 2.5 mmHg using the Omron M6, and -1.1 ± 4.8 and -0.9 ± 4.3 mmHg using the Omron R2.

Conclusion: Readings from the Omron M2, Omron M3, Omron M6, and Omron R2, differing by less than 5, 10, and 15 mmHg, fulfill the ESH international protocol and its 2010 revision requirements. Therefore, each of these four devices can be used by patients for SBPM.

Citing Articles

A Multi-Component Intervention to Improve Therapeutic Adherence in Uncontrolled Hypertensive Patients Within the Primary Care Level: A Before-and-After Study.

Gulayin P, Gutierrez L, Pinto D, Fontana S, Avila M, Gomez W High Blood Press Cardiovasc Prev. 2024; 31(3):271-278.

PMID: 38717676 DOI: 10.1007/s40292-024-00645-1.


Hemodynamic Bedside Monitoring Instrument with Pressure and Optical Sensors: Validation and Modality Comparison.

Kaisti M, Panula T, Sirkia J, Pankaala M, Koivisto T, Niiranen T Adv Sci (Weinh). 2024; 11(24):e2307718.

PMID: 38647263 PMC: 11200005. DOI: 10.1002/advs.202307718.


The role of supervision and motivation during exercise on physical and mental health in older adults: a study protocol for a randomized controlled trial (PRO-Training project).

Gomez-Redondo P, Valenzuela P, Martinez-de-Quel O, Sanchez-Martin C, Cerezo-Arroyo M, Moreno-Manzanaro D BMC Geriatr. 2024; 24(1):274.

PMID: 38509514 PMC: 10953175. DOI: 10.1186/s12877-024-04868-8.


Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring.

Panula T, Sirkia J, Koivisto T, Pankaala M, Niiranen T, Kantola I iScience. 2023; 26(11):108295.

PMID: 38026187 PMC: 10665806. DOI: 10.1016/j.isci.2023.108295.


Effects of Controlled Supramaximal High-Intensity Interval Training on Cardiorespiratory Fitness and Global Cognitive Function in Older Adults: The Umeå HIT Study-A Randomized Controlled Trial.

Simonsson E, Levik Sandstrom S, Hedlund M, Holmberg H, Johansson B, Lindelof N J Gerontol A Biol Sci Med Sci. 2023; 78(9):1581-1590.

PMID: 36972981 PMC: 10460559. DOI: 10.1093/gerona/glad070.


References
1.
OBrien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R . European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010; 15(1):23-38. DOI: 10.1097/MBP.0b013e3283360e98. View

2.
OBrien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J . Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit. 2002; 7(1):3-17. DOI: 10.1097/00126097-200202000-00002. View

3.
Mancia G, De Backer G, Dominiczak A, Cifkova R, Fagard R, Germano G . 2007 Guidelines for the Management of Arterial Hypertension: The Task Force for the Management of Arterial Hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). J Hypertens. 2007; 25(6):1105-87. DOI: 10.1097/HJH.0b013e3281fc975a. View

4.
OBrien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T . Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens. 2005; 23(4):697-701. DOI: 10.1097/01.hjh.0000163132.84890.c4. View

5.
OBrien E, Petrie J, Littler W, de Swiet M, Padfield P, Altman D . An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993; 11(6):677-9. DOI: 10.1097/00004872-199306000-00013. View