Development of a Prototype Gantry System for Preclinical X-ray Phase-contrast Computed Tomography
Overview
Authors
Affiliations
Purpose: To explore the potential of grating-based x-ray phase-contrast imaging for clinical applications, a first compact gantry system was developed. It is designed such that it can be implemented into an in-vivo small-animal phase-contrast computed tomography (PC-CT) scanner. The purpose of the present study is to assess the accuracy and quantitativeness of the described gantry in both absorption and phase-contrast.
Methods: A phantom, containing six chemically well-defined liquids, was constructed. A tomography scan with cone-beam reconstruction of this phantom was performed yielding the spatial distribution of the linear attenuation coefficient μ and decrement δ of the complex refractive index. Theoretical values of μ and δ were calculated for each liquid from tabulated data and compared with the experimentally measured values. Additionally, a color-fused image representation is proposed to display the complementary absorption and phase-contrast information in a single image.
Results: Experimental and calculated data of the phantom agree well confirming the quantitativeness and accuracy of the reconstructed spatial distributions of μ and δ. The proposed color-fused image representation, which combines the complementary absorption and phase information, considerably helps in distinguishing the individual substances.
Conclusions: The concept of grating-based phase-contrast computed tomography (CT) can be implemented into a compact, cone-beam geometry gantry setup. The authors believe that this work represents an important milestone in translating phase-contrast x-ray imaging from previous proof-of-principle experiments to first preclinical biomedical imaging applications on small-animal models.
Umkehrer S, Morrone C, Dinkel J, Aigner L, Reiser M, Herzen J Sci Rep. 2020; 10(1):18444.
PMID: 33116193 PMC: 7595203. DOI: 10.1038/s41598-020-75185-4.
Wu Z, Gao K, Wang Z, Wei C, Wali F, Zan G J Synchrotron Radiat. 2018; 25(Pt 4):1222-1228.
PMID: 29979185 PMC: 6038613. DOI: 10.1107/S1600577518008019.
X-Ray Dark-field Imaging to Depict Acute Lung Inflammation in Mice.
Hellbach K, Meinel F, Conlon T, Willer K, Yaroshenko A, Velroyen A Sci Rep. 2018; 8(1):2096.
PMID: 29391514 PMC: 5794739. DOI: 10.1038/s41598-018-20193-8.
Yang Q, Cong W, Wang G Proc SPIE Int Soc Opt Eng. 2017; 9967.
PMID: 28579667 PMC: 5453730. DOI: 10.1117/12.2237818.
X-ray Dark-field Radiography - In-Vivo Diagnosis of Lung Cancer in Mice.
Scherer K, Yaroshenko A, Bolukbas D, Gromann L, Hellbach K, Meinel F Sci Rep. 2017; 7(1):402.
PMID: 28341830 PMC: 5428469. DOI: 10.1038/s41598-017-00489-x.