Validation of a Photography-based Goniometry Method for Measuring Joint Range of Motion

Overview
Date 2011 Oct 11
PMID 21983191
Citations 30
Authors
Affiliations
Soon will be listed here.
Abstract

Background: A critical component of evaluating the outcomes after surgery to restore lost elbow motion is the range of motion (ROM) of the elbow. This study examined if digital photography-based goniometry is as accurate and reliable as clinical goniometry for measuring elbow ROM.

Materials And Methods: Instrument validity and reliability for photography-based goniometry were evaluated for a consecutive series of 50 elbow contractures by 4 observers with different levels of elbow experience. Goniometric ROM measurements were taken with the elbows in full extension and full flexion directly in the clinic (once) and from digital photographs (twice in a blinded random manner).

Results: Instrument validity for photography-based goniometry was extremely high (intraclass correlation coefficient: extension = 0.98, flexion = 0.96). For extension and flexion measurements by the expert surgeon, systematic error was negligible (0° and 1°, respectively). Limits of agreement were 7° (95% confidence interval [CI], 5° to 9°) and -7° (95% CI, -5° to -9°) for extension and 8° (95% CI, 6° to 10°) and -7° (95% CI, -5° to -9°) for flexion. Interobserver reliability for photography-based goniometry was better than that for clinical goniometry. The least experienced observer's photographic goniometry measurements were closer to the reference measurements than the clinical goniometry measurements.

Conclusions: Photography-based goniometry is accurate and reliable for measuring elbow ROM. The photography-based method relied less on observer expertise than clinical goniometry. This validates an objective measure of patient outcome without requiring doctor-patient contact at a tertiary care center, where most contracture surgeries are done.

Citing Articles

Validity Analysis of Monocular Human Pose Estimation Models Interfaced with a Mobile Application for Assessing Upper Limb Range of Motion.

Moreira R, Teixeira S, Fialho R, Miranda A, Lima L, Carvalho M Sensors (Basel). 2025; 24(24.

PMID: 39771719 PMC: 11679233. DOI: 10.3390/s24247983.


Long-Term Results of Suture-Button Suspensionplasty in the Treatment of Thumb Carpometacarpal Arthritis: A Minimum 10-Year Follow-Up.

Lachnish J, Titan A, Sen S, Yao J J Hand Surg Glob Online. 2024; 6(2):206-211.

PMID: 38903836 PMC: 11185894. DOI: 10.1016/j.jhsg.2023.12.002.


Remote visual estimation of shoulder range of motion has generally high interobserver reliability but limited accuracy.

Hwang S, Ardebol J, Ghayyad K, Pak T, Bonadiman J, Denard P JSES Int. 2023; 7(6):2528-2533.

PMID: 37969522 PMC: 10638579. DOI: 10.1016/j.jseint.2023.07.002.


Prosthetic resurfacing of engaging posterior capitellar defects in recurrent posterolateral rotatory instability of the elbow.

Rotman D, Rojas Lievano J, ODriscoll S Clin Shoulder Elb. 2023; 26(3):287-295.

PMID: 37652744 PMC: 10497932. DOI: 10.5397/cise.2022.01424.


The Intra- and Inter-Rater Reliability of a Variety of Testing Methods to Measure Shoulder Range of Motion, Hand-behind-Back and External Rotation Strength in Healthy Participants.

Beshara P, Davidson I, Pelletier M, Walsh W Int J Environ Res Public Health. 2022; 19(21).

PMID: 36361321 PMC: 9653808. DOI: 10.3390/ijerph192114442.