Precision Grid and Hand Motion for Accurate Needle Insertion in Brachytherapy

Overview
Journal Med Phys
Specialty Biophysics
Date 2011 Sep 21
PMID 21928648
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: In prostate brachytherapy, a grid is used to guide a needle tip toward a preplanned location within the tissue. During insertion, the needle deflects en route resulting in target misplacement. In this paper, 18-gauge needle insertion experiments into phantom were performed to test effects of three parameters, which include the clearance between the grid hole and needle, the thickness of the grid, and the needle insertion speed. Measurement apparatus that consisted of two datum surfaces and digital depth gauge was developed to quantify needle deflections.

Methods: The gauge repeatability and reproducibility (GR&R) test was performed on the measurement apparatus, and it proved to be capable of measuring a 2 mm tolerance from the target. Replicated experiments were performed on a 2(3) factorial design (three parameters at two levels) and analysis included averages and standard deviation along with an analysis of variance (ANOVA) to find significant single and two-way interaction factors.

Results: Results showed that grid with tight clearance hole and slow needle speed increased precision and accuracy of needle insertion. The tight grid was vital to enhance precision and accuracy of needle insertion for both slow and fast insertion speed; additionally, at slow speed the tight, thick grid improved needle precision and accuracy.

Conclusions: In summary, the tight grid is important, regardless of speed. The grid design, which shows the capability to reduce the needle deflection in brachytherapy procedures, can potentially be implemented in the brachytherapy procedure.

Citing Articles

Variables Affecting the Histological Quality of Prostate Core Biopsy Samples in Cancer Detection.

Elijaha I, Samsam A, Rafiei A, Patel T, Song S J Cancer Biol Res. 2024; 11(1).

PMID: 39691738 PMC: 11649315.


Needle Guide Friction Reduction Utilizing Contact Surface Structure.

Elijah I, Song S SM Radiol J. 2024; 7(1).

PMID: 39669008 PMC: 11634242.


IRE made easy: introducing the robotic grid system for multiple parallel needle insertion in irreversible electroporation treatment.

Wardhana G, Futterer J, Abayazid M Int J Comput Assist Radiol Surg. 2024; 19(8):1517-1526.

PMID: 38896406 PMC: 11329412. DOI: 10.1007/s11548-024-03216-w.


Percutaneous MR-guided whole-gland prostate cancer cryoablation: safety considerations and oncologic results in 30 consecutive patients.

De Marini P, Cazzato R, Garnon J, Tricard T, Koch G, Tsoumakidou G Br J Radiol. 2019; 92(1097):20180965.

PMID: 30845821 PMC: 6580912. DOI: 10.1259/bjr.20180965.


Surface markers for guiding cylindrical diffuser fiber insertion in interstitial photodynamic therapy of head and neck cancer.

Oakley E, Bellnier D, Hutson A, Wrazen B, Arshad H, Quon H Lasers Surg Med. 2017; 49(6):599-608.

PMID: 28185275 PMC: 5513772. DOI: 10.1002/lsm.22644.