Development of a Low-Cost SIA-Based Analyser for Water Samples

Overview
Date 2011 Jul 26
PMID 21785573
Authors
Affiliations
Soon will be listed here.
Abstract

An automated multiparametric water analyser was developed and evaluated. The system was based on Sequential Injection Analysis and featured a photometric detection system comprising a tricolour RGB LED source and a photodiode. A program compiled in Visual Basic was used to control the SIA flow system, the LEDs, and the data acquisition and processing. The program loads and executes methods written in ASCII and stored as text files. The system was capable of handling up to four methods simultaneously. When used to carry out methods based on the APHA standard methods, the figures of merit obtained were considered satisfactory for the purpose. The total cost was under US $4600. It was concluded that the analyser is appropriate for routine use and has potential for an increased number of simultaneous methods and for enhanced capabilities if new versions of the software are developed.

References
1.
Lyddy-Meaney A, Ellis P, Worsfold P, Butler E, McKelvie I . A compact flow injection analysis system for surface mapping of phosphate in marine waters. Talanta. 2008; 58(6):1043-53. DOI: 10.1016/s0039-9140(02)00428-9. View

2.
Miro M, Estela J, Cerda V . Application of flowing-stream techniques to water analysis Part II. General quality parameters and anionic compounds: halogenated, sulphur and metalloid species. Talanta. 2008; 62(1):1-15. DOI: 10.1016/S0039-9140(03)00415-6. View

3.
Dasgupta P, Bellamy H, Liu H, Lopez J, Loree E, Morris K . Light emitting diode based flow-through optical absorption detectors. Talanta. 1993; 40(1):53-74. DOI: 10.1016/0039-9140(93)80142-e. View

4.
Suzuki Y, Aruga T, Kuwahara H, Kitamura M, Kuwabara T, Kawakubo S . A simple and portable colorimeter using a red-green-blue light-emitting diode and its application to the on-site determination of nitrite and iron in river-water. Anal Sci. 2004; 20(6):975-7. DOI: 10.2116/analsci.20.975. View

5.
Gros N . Spectrometer with microreaction chamber and tri-colour light emitting diode as a light source. Talanta. 2008; 62(1):143-50. DOI: 10.1016/S0039-9140(03)00420-X. View