» Articles » PMID: 21782090

Objective Optical Assessment of Tear-film Quality Dynamics in Normal and Mildly Symptomatic Dry Eyes

Overview
Specialty Ophthalmology
Date 2011 Jul 26
PMID 21782090
Citations 41
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate and compare the tear-film dynamics in normal eyes and in eyes with mild dry-eye symptoms using a new noninvasive optical method based on a double-pass instrument.

Setting: Hospital Universitario Virgen de la Arrixaca, Murcia, Spain.

Design: Evaluation of diagnostic test or technology.

Methods: Dynamic recording of double-pass retinal images during unforced tear-film breakup was performed in eyes with mild dry-eye symptoms (study group) and in an asymptomatic control group. Series of consecutive retinal images were recorded every 0.5 seconds while the patient avoided blinking. Measurements were performed under low-light conditions to naturally increase pupil diameter and maximize the method's sensitivity. Additional clinical tests were performed for comparison and included tear breakup time (TBUT), Schirmer I tests, and a normalized questionnaire (McMonnies). From the retinal images, a quality metric, the intensity distribution index, was calculated. An objective TBUT value was estimated in each eye when the intensity distribution index surpassed a defined threshold value compared with the initial baseline.

Results: The study group comprised 20 eyes and the control group, 18 eyes. Symptomatic dry eyes had a typical exponential increase in the intensity distribution index with time. The objective TBUT values in the study group were comparable to the clinical TBUT estimates.

Conclusions: The new objective optical method to evaluate the quality and stability of the tear film was sensitive in detecting mild symptoms of dry eye and differentiating from normal cases. The procedure may allow early detection and follow-up of patients' tear film-related complaints.

Citing Articles

How ptosis affects the visual quality: an overview of visual quality impairments and contributing factors in ptotic eyes.

Shen J, Cui H Int Ophthalmol. 2025; 45(1):88.

PMID: 40085308 DOI: 10.1007/s10792-025-03468-8.


Compact and aberration effects-shielded objective intraocular scatter measurement system.

Zhao J, Zhang Z, Yang Y, Zhang H, Chen H, Wang S Biomed Opt Express. 2025; 16(2):669-678.

PMID: 39958848 PMC: 11828433. DOI: 10.1364/BOE.545245.


Biometry study of foveal isoplanatic patch variation for adaptive optics retinal imaging.

Huang X, Hargrave A, Bentley J, Dubra A Biomed Opt Express. 2024; 15(10):5674-5690.

PMID: 39421787 PMC: 11482173. DOI: 10.1364/BOE.536645.


Comparison of objective visual quality between SMILE and FS-LASIK in moderate-to-high myopia.

Zou H, Wei X, Li L, Wei D, Mao H, Huang Y Front Med (Lausanne). 2024; 11:1408516.

PMID: 39359915 PMC: 11446171. DOI: 10.3389/fmed.2024.1408516.


Effect of Lipiflow (Thermal Pulsation) on Ocular Surface Disease Management After Cataract Surgery.

Vasudevan B, Helmuth K, Fintelmann R Clin Ophthalmol. 2024; 18:2239-2252.

PMID: 39139652 PMC: 11319095. DOI: 10.2147/OPTH.S459472.