Defining a Safety Margin for Labral Suture Anchor Insertion Using the Acetabular Rim Angle

Overview
Journal Am J Sports Med
Publisher Sage Publications
Specialty Orthopedics
Date 2011 Jun 29
PMID 21709040
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Suture anchors are commonly used to reattach a torn labrum to the acetabular rim. The acetabular rim anatomy is not uniform, and the safety margin for inserting suture anchors is unknown. The acetabular rim angle is an anatomic measurement that is indicative of the safety margin for inserting suture anchors.

Purpose: To investigate the acetabular rim angle as a function of clock position, to evaluate the effect of drill depth on the acetabular rim angle, and to evaluate the effect of rim trimming on the acetabular rim angle.

Study Design: Descriptive laboratory study.

Methods: Three-dimensional acetabular models were reconstructed from computed tomography scans of 20 nonpaired cadaveric hip specimens, and the acetabular rim angle, which quantifies the angle between the subchondral margin and the outer cortex of the acetabulum, was measured from the 8- to 4-o'clock positions. At each position, the acetabular rim angle was measured for 5 drill depths (10, 12.5, 15, 20, and 25 mm) to simulate different lengths of suture anchors or drill bit depths on the acetabular rim angle. To simulate rim trimming, the acetabular rim angle was measured at the points that would become the suture anchor insertion points after 2.5- and 5-mm rim trimming.

Results: Clock position, drill depth, and rim trimming all had significant effects on the acetabular rim angle (P < .0001). The acetabular rim angle was largest at the 2-o'clock and smallest at the 3-o'clock position. Greater drill depths provided smaller acetabular rim angles, whereas rim trimming provided larger acetabular rim angles.

Conclusion: The acetabular rim angle varied significantly as a function of the location on the acetabular rim. A shorter drill depth and a greater amount of rim trimming provided a larger acetabular rim angle.

Clinical Relevance: Surgeons should be aware of the acetabular rim variations, especially in the anterosuperior quadrant, as well as the effects of drill depth and rim trimming, when selecting the optimal insertion angle for suture anchor placement to avoid articular cartilage penetration. The acetabular safety angle was smallest at the 3-o'clock position. Therefore, extra care must be taken when drilling or inserting anchors around the 3-o'clock position.

Citing Articles

Portal Selection for Suture Anchor Placement During Hip Arthroscopic Labral Repair: A Study Based on 3-Dimensional Model Reconstruction.

Chen Q, Zou J, Wang F, Qiao K, Li H, Zhang W Orthop J Sports Med. 2023; 11(8):23259671231189729.

PMID: 37655241 PMC: 10467395. DOI: 10.1177/23259671231189729.


Bone Volumes and Trajectory Angles for Acetabular Anchor Placement Can Be Optimized.

Di Loreto R, Getgood A, Degen R, Burkhart T Arthrosc Sports Med Rehabil. 2022; 4(2):e447-e452.

PMID: 35494283 PMC: 9042773. DOI: 10.1016/j.asmr.2021.10.022.


Perpendicular drill bit alignment provides a practical guidance to determine the appropriate suture anchor insertion angle during acetabular labral repair.

Ismailoglu A, Ozdogmus O, Karaman M, Kayaalp A, Kocaoglu B J Hip Preserv Surg. 2022; 8(2):185-191.

PMID: 35145716 PMC: 8825775. DOI: 10.1093/jhps/hnab055.


Fluoroscopy-Guided Suture Anchor Placement Yields Excellent Accuracy for Arthroscopic Acetabular Labral Repair: A Cadaveric Study.

Herickhoff P, Widner M, Mascoe J, Sebastianelli W Arthrosc Sports Med Rehabil. 2021; 3(5):e1505-e1511.

PMID: 34712987 PMC: 8527325. DOI: 10.1016/j.asmr.2021.07.012.


Allocation of Anchors During Labral Repair: A Multicenter Cohort Analysis of Labral Treatment in Hip Arthroscopy.

McGovern R, Christoforetti J, Kivlan B, Nho S, Wolff A, Salvo J Orthop J Sports Med. 2021; 9(2):2325967120981983.

PMID: 33681399 PMC: 7897831. DOI: 10.1177/2325967120981983.