MR Elastography of Human Lung Parenchyma: Technical Development, Theoretical Modeling and in Vivo Validation

Overview
Date 2011 May 19
PMID 21591003
Citations 32
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To develop a novel MR-based method for visualizing the elastic properties of human lung parenchyma in vivo and to evaluate the ability of this method to resolve differences in parenchymal stiffness at different respiration states in healthy volunteers.

Materials And Methods: A spin-echo MR Elastography (MRE) pulse sequence was developed to provide both high shear wave motion sensitivity and short TE for improved visualization of lung parenchyma. The improved motion sensitivity of this approach was modeled and tested with phantom experiments. In vivo testing was then performed on 10 healthy volunteers at the respiratory states of residual volume (RV) and total lung capacity (TLC).

Results: Shear wave propagation was visualized within the lungs of all volunteers and was processed to provide parenchymal shear stiffness maps for all 10 subjects. Density corrected stiffness values at TLC (1.83 ± 0.22 kPa) were higher than those at the RV (1.14 ± 0.14 kPa) with the difference being statistically significant (P < 0.0001).

Conclusion: (1)H-based MR elastography can noninvasively measure the shear stiffness of human lung parenchyma in vivo and can quantitate the change in shear stiffness due to respiration. The values obtained were consistent with previously reported in vitro assessments of cadaver lungs. Further work is required to increase the flexibility of the current acquisition and to investigate the clinical potential of lung MRE.

Citing Articles

Imaging the Lung in ARDS: A Primer.

Kaczka D Respir Care. 2024; 69(8):1011-1024.

PMID: 39048146 PMC: 11298232. DOI: 10.4187/respcare.12061.


Current evidence for lung ultrasound elastography in the field of pneumology: a systematic review.

Vargas-Ursua F, Ramos-Hernandez C, Pazos-Area L, Fernandez-Granda I, Rodriguez-Otero I, Gomez-Corredoira E ERJ Open Res. 2024; 10(4).

PMID: 39010887 PMC: 11247369. DOI: 10.1183/23120541.00081-2024.


Magnetic Resonance Elastography and Computational Modeling Identify Heterogeneous Lung Biomechanical Properties during Cystic Fibrosis.

Cho Y, Fakhouri F, Ballinger M, Englert J, Hayes D, Kolipaka A Res Sq. 2024; .

PMID: 38562870 PMC: 10984019. DOI: 10.21203/rs.3.rs-4125891/v1.


A Versatile MR Elastography Research Tool with a Modified Motion Signal-to-noise Ratio Approach.

Ito D, Habe T, Numano T, Okuda S, Soga S, Jinzaki M Magn Reson Med Sci. 2023; 23(4):417-427.

PMID: 37045750 PMC: 11447463. DOI: 10.2463/mrms.mp.2022-0149.


Quantitative Imaging Metrics for the Assessment of Pulmonary Pathophysiology: An Official American Thoracic Society and Fleischner Society Joint Workshop Report.

Hsia C, Bates J, Driehuys B, Fain S, Goldin J, Hoffman E Ann Am Thorac Soc. 2023; 20(2):161-195.

PMID: 36723475 PMC: 9989862. DOI: 10.1513/AnnalsATS.202211-915ST.


References
1.
Yin M, Talwalkar J, Glaser K, Manduca A, Grimm R, Rossman P . Assessment of hepatic fibrosis with magnetic resonance elastography. Clin Gastroenterol Hepatol. 2007; 5(10):1207-1213.e2. PMC: 2276978. DOI: 10.1016/j.cgh.2007.06.012. View

2.
Ito S, Ingenito E, Brewer K, Black L, Parameswaran H, Lutchen K . Mechanics, nonlinearity, and failure strength of lung tissue in a mouse model of emphysema: possible role of collagen remodeling. J Appl Physiol (1985). 2004; 98(2):503-11. DOI: 10.1152/japplphysiol.00590.2004. View

3.
Kononov S, Brewer K, Sakai H, Cavalcante F, Sabayanagam C, Ingenito E . Roles of mechanical forces and collagen failure in the development of elastase-induced emphysema. Am J Respir Crit Care Med. 2001; 164(10 Pt 1):1920-6. DOI: 10.1164/ajrccm.164.10.2101083. View

4.
Mariappan Y, Rossman P, Glaser K, Manduca A, Ehman R . Magnetic resonance elastography with a phased-array acoustic driver system. Magn Reson Med. 2009; 61(3):678-85. PMC: 2929704. DOI: 10.1002/mrm.21885. View

5.
Robert B, Sinkus R, Gennisson J, Fink M . Application of DENSE-MR-elastography to the human heart. Magn Reson Med. 2009; 62(5):1155-63. DOI: 10.1002/mrm.22124. View