» Articles » PMID: 21576270

Comparative Analysis of PCR-electrospray Ionization/mass Spectrometry (MS) and MALDI-TOF/MS for the Identification of Bacteria and Yeast from Positive Blood Culture Bottles

Overview
Journal Clin Chem
Specialty Biochemistry
Date 2011 May 18
PMID 21576270
Citations 41
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Emerging technologies for rapid identification of microbes demonstrate a shift from traditional biochemical and molecular testing algorithms toward methods using mass spectrometry (MS) for the semiquantitative analysis of microbial proteins and genetic elements. This study was performed to assess the diagnostic accuracy of 2 such technologies, PCR-electrospray ionization (ESI)/MS and MALDI-TOF/MS, with respect to phenotypic and biochemical profiling as a reference standard method. A positive challenge set of blood culture bottles was used to compare PCR-ESI/MS and MALDI-TOF/MS performance on a matched set of samples.

Methods: We performed characterization of bloodstream infections from blood cultures using the Ibis T5000 PCR-ESI/MS and the Bruker MALDI Biotyper 2.0 (MALDI-TOF/MS) platforms for microbial identification. Diagnostic accuracy was determined by independent comparison of each method to phenotypic and biochemical characterization with Vitek2 analysis as the reference standard identification.

Results: The diagnostic accuracy, represented as positive agreement, at the genus level was 0.965 (0.930-0.984) for PCR-ESI/MS and 0.969 (0.935-0.987) for MALDI-TOF/MS, and at the species level was 0.952 (0.912-0.974) with PCR-ESI/MS and 0.943 (0.902-0.968) for MALDI-TOF/MS. No statistically significant difference was found between PCR-ESI/MS and MALDI-TOF/MS in the ability to rapidly identify microorganisms isolated from blood culture.

Conclusions: Our results demonstrate that PCR-ESI/MS and MALDI-TOF/MS are equivalent in their ability to characterize bloodstream infections with respect to the reference standard, and highlight key differences in the methods that allow for each method to have a unique niche as a tool for rapid identification of microbes in blood cultures.

Citing Articles

Utility of Matrix-Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry in Periprosthetic Joint Infection Diagnosis.

Sandu E, Cursaru A, Iordache S, Serban B, Costache M, Cirstoiu C Cureus. 2024; 16(10):e70650.

PMID: 39483552 PMC: 11527459. DOI: 10.7759/cureus.70650.


Rapid Detection of Urinary Tract Infection in 10 min by Tracking Multiple Phenotypic Features in a 30 s Large-Volume Scattering Video of Urine Microscopy.

Zhang F, Mo M, Jiang J, Zhou X, McBride M, Yang Y ACS Sens. 2022; 7(8):2262-2272.

PMID: 35930733 PMC: 9465977. DOI: 10.1021/acssensors.2c00788.


Total Laboratory Automation for Rapid Detection and Identification of Microorganisms and Their Antimicrobial Resistance Profiles.

Cherkaoui A, Schrenzel J Front Cell Infect Microbiol. 2022; 12:807668.

PMID: 35186794 PMC: 8851030. DOI: 10.3389/fcimb.2022.807668.


Optimization of sepsis therapy based on patient-specific digital precision diagnostics using next generation sequencing (DigiSep-Trial)-study protocol for a randomized, controlled, interventional, open-label, multicenter trial.

Brenner T, Skarabis A, Stevens P, Axnick J, Haug P, Grumaz S Trials. 2021; 22(1):714.

PMID: 34663439 PMC: 8522064. DOI: 10.1186/s13063-021-05667-x.


Detection of Intrinsically Resistant Candida in Mixed Samples by MALDI TOF-MS and a Modified Naïve Bayesian Classifier.

Gong J, Shen C, Xiao M, Zhang H, Zhao F, Zhang J Molecules. 2021; 26(15).

PMID: 34361627 PMC: 8348127. DOI: 10.3390/molecules26154470.


References
1.
Sampath R, Hofstadler S, Blyn L, Eshoo M, Hall T, Massire C . Rapid identification of emerging pathogens: coronavirus. Emerg Infect Dis. 2005; 11(3):373-9. PMC: 3298233. DOI: 10.3201/eid1103.040629. View

2.
Peters R, van Agtmael M, Danner S, Savelkoul P, Vandenbroucke-Grauls C . New developments in the diagnosis of bloodstream infections. Lancet Infect Dis. 2004; 4(12):751-60. DOI: 10.1016/S1473-3099(04)01205-8. View

3.
La Scola B, Raoult D . Direct identification of bacteria in positive blood culture bottles by matrix-assisted laser desorption ionisation time-of-flight mass spectrometry. PLoS One. 2009; 4(11):e8041. PMC: 2777307. DOI: 10.1371/journal.pone.0008041. View

4.
Nagy E, Maier T, Urban E, Terhes G, Kostrzewa M . Species identification of clinical isolates of Bacteroides by matrix-assisted laser-desorption/ionization time-of-flight mass spectrometry. Clin Microbiol Infect. 2009; 15(8):796-802. DOI: 10.1111/j.1469-0691.2009.02788.x. View

5.
Raja S, Ching J, Xi L, Hughes S, Chang R, Wong W . Technology for automated, rapid, and quantitative PCR or reverse transcription-PCR clinical testing. Clin Chem. 2005; 51(5):882-90. DOI: 10.1373/clinchem.2004.046474. View